MEASUREMENT
13/10/2016 Version No. 001
57
Temperature Measurement
Background
The temperature measurement function uses the optional PT-100
temperature probe. The measured temperature is displayed on
the display. For more information on the optional PT-100 sensor,
There is only one range for the temperature function. However
the resistance measurement range can still be changed when in
the temperature function.
Note
The temperature measurement function is used in conjunction
with the Ohm measurement function. The two measurements
share the same display, so the Ohm readings stay on the display
even after the temperature measurement function is activated.
Thus when the Temperature function is selected, “Ohm+T” is
shown as the selected function.
1. Select the
Temperature
function
Press
TEMP
to enter the temperature measurement function.
Tempe Ohm
function indicator
(Ambient)
temperature
source
Resistance measurement
Ambient
temperature
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...
Страница 140: ...RM 804 RM 805 Instruction Manual English 138 13 10 2016 Version No 001 Example BINN LIM RES 1 Indicates a pass for bin1...