RM-804 & RM-805 Instruction Manual/English
156
13/10/2016 Version No. 001
FAST
Measurement speed is fast.
Example
SENS:SPE FAST
Sets measurement speed to the fast rate.
SENSe:REL:DATa
Set
Query
Description
Sets or returns the relative value for the relative function.
Syntax
Query Syntax
SENSe:REL:DATa <NRf>
SENSe:REL:DATa?
Parameter
<NRf>
0.0000~500.00
The unit will be auto set by the present
range.
Return parameter
<NR3>
±0.0000~5.1000E±X
Example
SENS:REL:DAT 490.32
Sets the relative function value to 490.32Ω.
SENS:REL:DAT?
>4.9032E+2
Returns the relative value (490.32Ω).
SENSe:REL:STATe
Set
Query
Description
Sets or returns the relative function state