MEASUREMENT
13/10/2016 Version No. 001
59
Temperature Compensation
Background
If the resistance of a DUT at a particular temperature is needed,
the compensation function can be used. This function can
simulate the resistance of a DUT at a desired temperature. If the
ambient temperature and the temperature coefficient of the DUT
are known, it is possible to determine the resistance of a DUT at
any temperature.
The Temperature Compensation works on the following formula:
)
(
1
0
0
0
t-t
α
R
R
t
t
t
Where:
R
t
= Measured resistance value (Ω)
R
t0
= Corrected resistance value (Ω)
T
0
= Inferred absolute temperature
t
0
= Corrected temperature (ºC)
t
= Current ambient temperature (ºC)
α
to
= Temperature coefficient of resistance at the correct
temperature.
0
0
|
1
t
T
|
a
to
1. Select the
Temperature
Compensation
Press
TC
to access the Temperature Compensation
function.