RM-804 & RM-805 Instruction Manual/English
182
13/10/2016 Version No. 001
parameter
<NR1>
1~8:bin1~bin8 state
9:bin out state
10:hi state
11:low state
12:pass state
13:fail state
Example
USER1:FIRS 12
Sets first operand of userdefine1 as pass state.
USERdefine<X>:LOGic
Set
Query
Description
Sets or returns operator for the selected user define pin.
Syntax
Query Syntax
USERdefine<X>:LOGic <NR1>
USERdefine<X>:LOGic?
Parameter/ Return
parameter
<X>
Userdefine pin 1~2
<NR1>
1:off(only judge first data)
2:logical and.
3:logical or.
Example
USER1:LOG 1
Sets the operator of userdefine1 to off. (I.e., only the first
operand determines the output of userdefine1.)
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...
Страница 140: ...RM 804 RM 805 Instruction Manual English 138 13 10 2016 Version No 001 Example BINN LIM RES 1 Indicates a pass for bin1...