RM-804 & RM-805 Instruction Manual/English
40
13/10/2016 Version No. 001
Real-time measurement
Dry-Circuit Measurement
Background
The Dry Circuit measurement function is used where the
maximum open-circuit voltage must be kept to a minimum for
applications such as measuring the contact resistance of
switches, relays and connectors. The RM-805 provides a
maximum of up to 20mV in this mode.
Note
Dry circuit testing is for switch and connector contact resistance.
Switch and connector contact resistance measurement is in
accordance with DIN IEC 512 and ASTM B539 which requires
that the open circuit voltage of the measuring device should not
exceed 20mV DC. Voltage at such low levels avoids the