RM-804 & RM-805 Instruction Manual/English
34
13/10/2016 Version No. 001
DC+
~ +6.5V
Open circuit
voltage
0V
V
t
Default drive signal.
DC-
~ -6.5V
Open circuit
voltage
0V
V
t
Negative drive
signal.
Pulse
~-6.5V
0V
V
t
~+6.5V
50ms
50ms
This mode can be used to
eliminate the thermoelectric EMF
formed on the contact between a
test lead and a DUT.
PWM
0V
V
t
~+6.5V
ON duty
This mode can be used to avoid
heating up the DUT and thus
avoid having the measurement
accuracy compromised on
temperature-sensitive DUTs.