RM-804 & RM-805 Instruction Manual/English
186
13/10/2016 Version No. 001
Example
*IDN?
>RS PRO,RM805,GXXXXXXXX,V1.00.
*OPC
Set
Query
Description
Sets or returns the operation complete bit (bit0) in SERS
(Standard Event Status Register) when all pending
operations are completed.
Syntax
Query Syntax
*OPC
*OPC?
Parameter
<None>
Return parameter
<NR1>
0:operation not complete
1:operation complete
Example
*OPC?
Returns 1.
*RST
Set
Query
Description
Recalls default panel setup.
Syntax
*RST
Parameter
<None>
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...