COMMAND OVERVIEW
13/10/2016 Version No. 001
139
Calculate Commands
CALCulate:COMPare:BEEPer
Set
Query
Description
Sets or returns the compare function beeper mode.
Syntax
Query Syntax
CALCulate:COMPare:BEEPer {OFF|PASS|FAIL}
CALCulate:COMPare:BEEPer?
Parameter/ Return
parameter
OFF
Turns the beeper off.
PASS
The beeper will sound on a pass test result.
FAIL
The beeper will sound on a fail test result.
Example
CALC:COMP:BEEP FAIL
Sets the beeper on when the test result is a fail.
CALCulate:COMPare:LIMit:LOWer
Set
Query
Description
Sets or returns the lower limit value for the compare function.
Syntax
Query Syntax
CALCulate:COMPare:LIMit:LOWer {<NRf>[,<String>]}
CALCulate:COMPare:LIMit:LOWer?
Parameter
<NRf>
000.0000~999.9999
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...
Страница 140: ...RM 804 RM 805 Instruction Manual English 138 13 10 2016 Version No 001 Example BINN LIM RES 1 Indicates a pass for bin1...