MEASUREMENT
13/10/2016 Version No. 001
41
breakdown of any oxides that may be present on the contacts. In
this mode the open circuit measuring voltage is limited <20mV,
while modes like DC+ or pulse mode can have an open circuit
measuring voltage as high as 6.5V.
The Dry Circuit function cannot be used with the Scan or Diode
functions. In addition, when the Dry Circuit function is turned on,
only 3 drive settings are available: DC+, DC- and Pulse.
Dry Limitations
When the Dry Circuit measurement function is turned on, the
measurement range is reduced. See the specifications for more
details.
Range
Dry Mode
Rate
50mΩ
✘
500mΩ
✔
Slow/Fast
5Ω
✔
Slow/Fast
50Ω
✔
Slow/Fast
500Ω
✘
5kΩ
✘
50kΩ
✘
500kΩ
✘
5MΩ
✘
1. Toggle Dry
mode on or off
Press the
Dry
key to toggle the dry circuit measurement
mode on or off.
The DRY function indicator will appear in the middle of the display