RM-804 & RM-805 Instruction Manual/English
70
13/10/2016 Version No. 001
avoid the effects of transient disturbances that are usually seen
when measuring reactive DUTs with a current source.
1. Select
Measure setting
From one of the main screens, press the
ESC
key so that the menu system at
the bottom of the display has focus.
Go to Meas. Setup and press Enter.
Go to Measure Delay and press Enter.
Meas. Setup
menu icon
Move
Select menu
or setting
Enter
2. Measure
Delay setting
appears
Use the arrow keys to turn Measure Delay on and set the delay
time. Press Enter to confirm the setting.
Measure
delay setting
Measure Delay*
OFF, ON: 000.000 ~ 100.000s
* When the set value is > 0.1s, the resolution is 0.1s. When the
set value is < 0.1S, the resolution is 1mS.
Note
Pressing ESC before pressing ENTER will exit the Measure
Delay settings.
Trigger Delay