COMMAND OVERVIEW
13/10/2016 Version No. 001
163
ON
Turn the average function on.
Example
SYST:AVER:STAT OFF
Turns the average function off.
SYSTem:BRIGhtness
Set
Query
Description
Sets or returns the brightness level.
Syntax
Query Syntax
SYSTem:BRIGhtness <NR1>
SYSTem:BRIGhtness?
Parameter/ Return
parameter
<NR1>
1(dim)~5(bright)
Example
SYST:BRIG 4
Turns the brightness level to 4.
SYSTem:ERRor
Query
Description
Returns the current system error, if any.
Query Syntax
SYSTem:ERRor?
Return parameter
<String>
Error number,”Error message”
Example
SYST:ERR?
>0,”No error”.
Indicates that there is no error message.