RM-804 & RM-805 Instruction Manual/English
60
13/10/2016 Version No. 001
mode
The temperature-compensated resistance measurement will
appear on the display.
Temperature
compensation
function indicator
Ambient
temperature
source
Extrapolated resistance
measurement at the desired
(
“correct”) temperature
Correct Temperature,
Temperature Coefficient settings
Ambient
temperature
2. Ambient
Temperature
The ambient temperature can be either measured with the PT-
100 sensor or be set manually.
If using the PT-100 sensor the Ambient temperature setting
should be turned off. If the PT-100 probe is not used, then the
ambient temperature needs to be manually set.
From the bottom menu, go to Meas. Setup > Ambient
Temperature and set the ambient temperature.
Range
Off, -50.0 ºC ~ 399.9ºC