RM-804 & RM-805 Instruction Manual/English
136
13/10/2016 Version No. 001
Parameter/
Return parameter
COMP
The display is set to compare mode.
COUNT
The display is set to count mode.
Example
BINN:LIM:DISP COMP
Sets the bin sorting function display mode to compare.
BINNing:LIMit:MODE
Set
Query
Description
Sets or returns the setting mode for upper and lower limits
(absolute or Δ%).
Syntax
Query Syntax
BINNing:LIMit:MODE {ABS|DPER}
BINNing:LIMit:MODE?
Parameter/
Return parameter
ABS
The test results are judged from absolute
values.
DPER
The test results are judged from a reference
value ± a percentage offset.
(delta percent)
Example
BINN:LIM:MODE DPER
Sets the mode to Δ%.
BINNing:LIMit:REFerence
Set
Query
Description
Sets or returns the limit reference value for the bin sorting
function.