RM-804 & RM-805 Instruction Manual/English
140
13/10/2016 Version No. 001
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter
<NR3>
000.0000~999.9999E±X
Example
CALC:COMP:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:COMP:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.
CALCulate:COMPare:LIMit:MODE
Set
Query
Description
Sets or returns the compare mode for the compare function.
Syntax
Query Syntax
CALCulate :COMPare:LIMit:MODE {ABS|DPER|PER}
CALCulate:COMPare:LIMit:MODE?
Parameter/
Return parameter
ABS
The test results are judged from absolute
values.
DPER
The test results are judged from a reference
value ± a percentage offset. (delta
percentage)
PER
The test results are displayed as a
percentage of the reference value.
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...
Страница 140: ...RM 804 RM 805 Instruction Manual English 138 13 10 2016 Version No 001 Example BINN LIM RES 1 Indicates a pass for bin1...