RM-804 & RM-805 Instruction Manual/English
152
13/10/2016 Version No. 001
Parameter
<NR1>
1~20
Example
MEM:SAV 1
Saves the settings to memory slot 1.
MEMory:STATe
Query
Description
Returns the status of all the memory slots.
Syntax
MEMory:STATe?
Return parameter
<String>
23 Characters composed of “N” or “F”,
where “N” indicates “Not used” and “F”
indicates “Full”.
Example
MEM:STAT?
> NFFNN-NNNNN-NNNNN-NNNNN
Indicates that memory slots 2 and 3 have data and that all
other memory slots are empty.
Содержание RM-804
Страница 1: ...Instruction Manual RM 804 RM 805 D C Milli Ohm Meter EN FR DE IT SP...
Страница 2: ......
Страница 69: ...MEASUREMENT 13 10 2016 Version No 001 67 DUT connection 4 wire Kelvin GUARD SOURCE SENSE SENSE SOURCE shielding...