IIT-2000/2010/5000/5010 Instruction Manual/English
196
05/08/2015 Version No. 001
When I press the START button the tester will not start testing?
The tester must first be in the READY status before a test can be started. Ensure
the tester displays READY before pressing the START button, page 72 (manual
test), 102(automatic test).
If “Double Action” is enabled, the START button must be pressed 0.5 seconds after
the STOP button is pressed, otherwise the tester will not start testing.
If “Interlock” is enabled, the interlock key must be inserted into the signal I/O port
on the rear before a test can be started. See page 127 for details.
Lastly, ensure that the Start Ctrl setting is correctly configured in the Common Utility
menu. For example, to enable the START button to start a test, ensure that the
Start Ctrl setting is set to FRONT PANEL. See page 117 for details.
The accuracy does not match the specification.
Make sure the tester is powered on for at least 30 minutes, 15°C~+35°C.
This is necessary to stabilize the unit to match the specification.
For more information, contact your local dealer or RS Components at www.rs-
components.com