IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
61
Note
The ARC setting range is directly related to the HI SET current
limit.
ACW: IIT-2000/2010
HI SET Limit
ARC Range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~042.0mA
001.0mA ~080.0mA
ACW: IIT-5010/5000
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~110.0mA
002.0mA ~200.0mA
DCW: IIT-2000/2010
HI SET Limit
ARC Range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~011.0mA
001.0mA ~020.0mA
DCW: IIT-5010/5000
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~021.0mA
002.0mA ~040.0mA