![RS PRO 917-6298 Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/rs-pro/917-6298/917-6298_instruction-manual_1496054042.webp)
IIT-2000/2010/5000/5010 Instruction Manual/English
40
05/08/2015 Version No. 001
HIGH
VOLTAGE
Terminal
RETURN
Terminal
GB Connection
Background
GB tests use the SENSE H/L and SOURCE H/L terminals with
the GTL-115 test leads.
GB Connection
Source H
DUT
IIT-2010/5010
Sense L
Sense H
Source L
Steps
1. Turn the power off on the safety tester.
2. Connect the Sense H lead to the SENSE H terminal.
3. Connect the Sense L lead to the SENSE L terminal.
4. Connect the Source H lead to the SOURCE H terminal.
5. Connect the Source L lead to the SOURCE L terminal.