IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
161
Example
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
MANU:ACW:REF
Set
Query
Description
Sets or returns the ACW reference value in mA. The test must
first be in ACW mode before this command can be used.
The ACW reference value must be less than the HI SET value.
The ACW reference value must use the same range as the HI
SET value.
Syntax
Query Syntax
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/
Return
parameter
<NR2> 0.000 ~ 041.9 (IIT-2000/2010)
0.000 ~ 109.9 (IIT-5010/5000)
Example
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.