IIT-2000/2010/5000/5010 Instruction Manual/English
174
05/08/2015 Version No. 001
Parameter/
Return
parameter
<NR2> 0.000 ~ 649.9
Example
MANU:GB:REF 100
Sets the GB reference to 100 mΩ.
MANU:GB:ZEROCHECK
Set
Query
Description
Performs the zero check function. The test must first be in GB
mode and in the Ready Status before this command can be used.
See page 83 for details on the ZERO function.
Syntax
Query Syntax
MANU:GB:ZEROCHECK {ON|OFF}
MANU:GB:ZEROCHECK?
Parameter/
Return
parameter
ON
Zero function is active.
OFF
Zero function is not active.
Example
MANU:GB:ZEROCHECK OFF
Activates the ZERO function.
MANU:UTILity:ARCMode
Set
Query
Description
Sets or returns the ARC mode status for the current test.
The ARC mode cannot be set for the IR and GB function.