IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
155
MANU:NAME
Set
Query
Description
Sets or returns the test name for the selected manual test. The
test must be in MANU mode before this command can be used.
Note only alphanumeric characters (A-Z, a-z, 0-9) and t
he “_”
underscore character can be used to set the MANU test name.
Syntax
Query Syntax
MANU:NAME <string>
MANU:NAME?
Parameter/
Return
parameter
<string> 10 character string. (first character must be a letter)
Example
MANU:NAME test1
Sets the manual test
name to “test1”.
MANU:INITial
Set
Query
Description
Loads the initial (default) settings for the selected MANU test
number. The initial settings that are loaded depend on the test
function (ACW, DCW, IR or GB).
Syntax
MANU:INITial
Initial Settings
Parameter
Function
ACW
DCW
IR
GB
REF#
0.000mA
0.000mA
0000MΩ
000.0mΩ
FREQ
60Hz
X
X
60Hz
Содержание 917-6298