![RS PRO 917-6298 Скачать руководство пользователя страница 80](http://html.mh-extra.com/html/rs-pro/917-6298/917-6298_instruction-manual_1496054080.webp)
IIT-2000/2010/5000/5010 Instruction Manual/English
78
05/08/2015 Version No. 001
Pressing the STOP button will return the tester
to the READY status.
STOP
Pressing the START button will restart the test.
START
Note
The buzzer will only sound if the Pass Sound is set to ON.
See page 113 for details.
The START button is disabled when the buzzer is beeping.
PASS Timing
Diagrams
The timing diagrams below show the ACW, DCW, IR and GB
timing for the START status, TEST status and PASS judgment.
ACW PASS
Timing
Discharge time
Output V
time
Initial time
START
TEST
PASS
RAMP
TEST TIME
(Approximately 150ms)