Error Messages
Description
TIME ERR
For ACW tests.
IIT-2000/2010:
TIME ERR is displayed w
hen HI SET ≥ 30.00mA~40.00mA
and if the RAMP time and the TEST TIME setting is >
240 seconds.
IIT-5010/5000:
TIME ERR is displayed when HI SET ≥ 80.00mA~100.0mA
and if the RAMP time and the TEST TIME setting is >
240 seconds.
OVER 50W (IIT-
2000/2010)
For DCW tests. OVER 50W is displayed if the HI SET
setting multiplied by the Voltage setting is greater than
50W (IIT-2000/2010 only).
OVER 100W
(IIT-5010/5000)
For DCW tests. OVER 100W is displayed if the HI SET
setting multiplied by the Voltage setting is greater than
100W (IIT-5010/5000 only).
I ERR
For ACW, DCW tests. Shown when the current is set too
high.
SHORT
Voltage is too low or there is no High Voltage output.
Indicates that the DUT could be shorted.
V ERR
For ACW, DCW tests. Indicates that an abnormal voltage
has been detected.
V = 0
For GB tests. Voltage is equal to 0. Check to see that the
SENSE H or SOURCE H is not open.