IIT-2000/2010/5000/5010 Instruction Manual/English
160
05/08/2015 Version No. 001
240 seconds when the HI SET limit is over 30mA (IIT-2000/2010)
or over 80mA (IIT-5010/5000). This applies to the ACW function
only.
In special MANU mode, the TIMER can be turned off.
Syntax
Query Syntax
MANU:ACW:TTIMe {<NR2>|OFF}
MANU:ACW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return
parameter
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
Example
MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.
MANU:ACW:FREQuency
Set
Query
Description
Sets or returns the ACW test frequency in Hz. The test must first
be in ACW mode before this command can be used.
Syntax
Query Syntax
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/
Return
parameter
50
50 Hz
60
60 Hz