IIT-2000/2010/5000/5010 Instruction Manual/English

 

05/08/2015 Version No. 001 

209

 

AC 

8

0mA≤I≤100m

A  

At least as long 

as the output 

time 

Maximum 240 seconds 

 

0.001m

A≤I<80

mA 

Not necessary  Continuous output possible 

DC 

0.001mA≤I≤20

mA 

Not necessary  Continuous output possible 

GB  

15A<I≤32A 

(IIT-5010) 

At least as long 

as the output 

time 

999.9 

 

3A≤I≤15A 

Not necessary  999.9 

NOTE: Output Time = Ramp Time + Test Time. 

Содержание 917-6298

Страница 1: ...Instruction Manual IIT 2000 2010 5000 5010 Electrical Safety Tester EN FR DE IT SP ...

Страница 2: ......

Страница 3: ...iew 10 Appearance 15 Set Up 24 OPERATION 32 Menu Tree 34 Test Lead Connection 39 ACW DCW IR and GB Manual Testing 42 Special MANU Test Mode 000 86 Automatic Tests 92 Common Utility Settings 112 EXTERNAL CONTROL 121 External Control Overview 122 REMOTE CONTROL 129 Interface Configuration 130 Command Syntax 136 Command List 139 Error Messages 194 FAQ 195 APPENDIX 197 ...

Страница 4: ...lish 2 05 08 2015 Version No 001 Fuse Replacement 197 Test Errors 198 IIT 2000 2010 5000 5010 Specifications 201 IIT 2000 Dimensions 210 IIT 2010 Dimensions 211 IIT 5010 Dimensions 212 IIT 5000 Dimensions 213 Declaration of Conformity 215 INDEX 216 ...

Страница 5: ...re your safety and to keep the instrument in the best possible condition Safety Symbols These safety symbols may appear in this manual or on the instrument WARNING Warning Identifies conditions or practices that could result in injury or loss of life CAUTION Caution Identifies conditions or practices that could result in damage to the instrument or to other properties DANGER High Voltage Attention...

Страница 6: ...gh handling that leads to damaging the instrument Do not discharge static electricity to the instrument Use only mating connectors not bare wires for the terminals Do not block the cooling fan opening Do not disassemble the IIT 2000 2010 5000 5010 unless you are qualified Measurement categories EN 61010 1 2010 specifies the measurement categories and their requirements as follows The IIT 2000 2010...

Страница 7: ... solution of mild detergent and water Do not spray any liquid Do not use chemicals containing harsh material such as benzene toluene xylene and acetone Operation Environment Location Indoor no direct sunlight dust free almost non conductive pollution Note below Relative Humidity 70 no condensation Altitude 2000m Temperature 0 C 40 C Pollution Degree EN 61010 1 2010 specifies the pollution degrees ...

Страница 8: ...to condensation which is expected In such conditions equipment is normally protected against exposure to direct sunlight precipitation and full wind pressure but neither temperature nor humidity is controlled Storage environment Location Indoor Temperature 10 C to 70 C Relative Humidity 85 no condensation Disposal Do not dispose this instrument as unsorted municipal waste Please use a separate col...

Страница 9: ...h is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the equipment or contact the supplier This cable appliance should be protected by a suitably rated and approved HB...

Страница 10: ...r in a nutshell including its main features and front rear panel introduction After going through the overview please read the safety considerations in the Set Up chapter IIT 2000 2010 5000 5010 Overview 10 Series lineup 10 Model Overview 11 Main Features 11 Accessories 12 Package Contents 14 Appearance 15 IIT 2000 5000 Front Panel 15 ...

Страница 11: ...No 001 9 IIT 2010 5010 Front Panel 15 IIT 2000 2010 Rear Panels 20 IIT 5000 Rear Panel 20 IIT 5010 Rear Panel 21 Set Up 24 Line Voltage Connection and Power Up 24 Installing the Optional GPIB Card 26 Workplace Precautions 27 Operating Precautions 28 Basic Safety Checks 31 ...

Страница 12: ... 6kVDC for DC withstanding voltage testing For the IIT 5010 5000 models the testing terminals are also mirrored on the rear panel for added safety and for more permanent safety testing environments They also include an innovative sweep function to view test results as a graph The IIT 2000 2010 5000 5010 can store up to 100 manual tests as well as run up to 16 manual tests sequentially as an automa...

Страница 13: ...015 Version No 001 11 Model Overview Model name ACW DCW IR GB Sweep IIT 2000 IIT 2010 IIT 5000 IIT 5010 Main Features Performance ACW 5kVAC DCW 6kVDC IR 50V 1000V 50V steps GB 3A 30A IIT 2010 3A 32A IIT 5010 The IIT 5010 also includes an extra 125V test point ...

Страница 14: ...ge and Ready indicators PWM output 90 efficiency increased reliability Interlock configurable Sweep Function Rear panel output IIT 5000 5010 only Interface Remote control start stop interface terminal RS232 USB interface for programming Optional GPIB interface for programming Signal I O port for pass fail test monitoring and start stop control interlock Accessories Standard Accessories Part number...

Страница 15: ...mote terminal male plug N A Interlock key Optional Accessories Part number Description GHT 205 High Voltage Test Probe GHT 113 High Voltage Test Pistol GTL 232 RS232C cable GTL 248 GPIB cable GTL 247 USB cable GRA 417 Rack Adapter Panel 19 4U IIT 2000 2010 5000 only Options Part number Description Opt 01 GPIB Interface GPIB module ...

Страница 16: ...ents single unit IIT 2000 2010 5000 5010 Quick Start guide User manual CD CTC Calibration Traceable Certificate Power cord x1 region dependent GHT 114 test leads x1 GTL 115 test leads x1 IIT 2010 5010 Remote terminal male plug Interlock key Note Keep the packaging including the box polystyrene foam and plastic envelopes should the need arise to return the unit to RS components ...

Страница 17: ... PASS FAIL indicators Directional keys Function keys HIGH VOLTAGE indicator RETURN terminal REMOTE terminal STOP button START button POWER button Display Scroll wheel Configuration keys READY indicator TEST indicator HIGH VOLTAGE output terminal IIT 2010 5010 Front Panel REMOTE MANU AUTO EDIT SAVE UTILITY ESC PAGE HIGH VOLTAGE CAUTION 5 0 kVAC MAX 6 0 kVDC MAX GB Rx SENSE H SOURCE H SENSE L SOURCE...

Страница 18: ...utomatic test ESC key ESC The ESC key is used to exit out of a menu or cancel a setting PAGE key PAGE The PAGE key is used to view automatic test information and test results Directional arrow keys The directional arrow keys are used to navigate menus and parameter settings READY indicator READY The READY indicator is lit when the tester is ready to begin testing The STOP button is used to put the...

Страница 19: ...OLTAGE terminal output is used for outputting the testing voltage The terminal is recessed for safety This terminal is used in conjunction with the RETURN terminal WARNING USE EXTREME CAUTION Do not touch the HIGH VOLTAGE terminal during testing RETURN terminal All models except IIT 2000 5000 RETURN The RETURN terminal is used for IR DCW and ACW tests RETURN SENSE and SOURCE terminals IIT 2010 501...

Страница 20: ...E key EDIT SAVE Used to start editing MANU AUTO tests as well as save settings and parameters MANU AUTO key MANU AUTO The MANU AUTO key is used to select manual tests MANU or automatic tests AUTO REMOTE terminal REMOTE The REMOTE terminal is used to connect to a remote controller STOP button STOP The STOP button is used to stop cancel tests The STOP button will also put the safety tester in the RE...

Страница 21: ... used to start tests The START button can be used to start tests when the tester is in the READY status Pressing the START button will put the tester in the TEST status POWER switch POWER Turns the power on The safety tester will always start up with the last test setting from when the instrument was last powered down ...

Страница 22: ...32 SER NO LB ENSURETHEPOWERISREMOVEDFROM THEINSTRUMENTBEFOREREPLACINGTHEFUSE GPIB T2 5A 250V IIT 5000 Rear Panel TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR MUST BE ONLY WITH SPECIFIED TYPE AND RATED FUSE NO OPERATOR SERVICEABLE COMPONENTS INSIDE DO NOT REMOVE COVERS REFER SERVICING TO FOR CONTINUED FIRE PROTECTION REPLACE CONNECTED TO GROUND QUALIFIED PERSONNEL WARNING A...

Страница 23: ... 60Hz 250V 250V T 10A FUSE T 6 3A MAX 1000VA GND ENSURE THE POWER IS REMOVED FROM THE INSTRUMENT BEFORE REPLACING THE FUSE POWER SIGNAL I O RS232 GPIB GB Rx SENSE H SOURCE H SENSE L SOURCE L RETURN HI POT IR HIGH VOLTAGE CAUTION MAX 6 0 kVDC 5 0 kVAC SER NO LB High voltage terminal Signal I O RS232 USB port Fan vents Fuse selector Line voltage Optional GPIB port GND Return Sense and Source termina...

Страница 24: ...GND GND Connect the GND ground terminal to the earth ground Line voltage input 220 230 100 120 Line voltage input 100 120 220 230VAC 10 Line voltage fuse 220 230 100 120 Line voltage selector and fuse IIT 2000 2010 100V 120V T5A 250V 220V 230V T2 5A 250V IIT 5010 5000 100V 120V T10A 250V 220V 230V T6 3A 250V Optional GPIB port GPIB Optional GPIB interface for remote control ...

Страница 25: ... outputting the testing voltage WARNING USE EXTREME CAUTION Do not touch the HIGH VOLTAGE terminal during testing RETURN terminal IIT 5000 RETURN The RETURN terminal is used for IR DCW and ACW tests RETURN SENSE and SOURCE terminals IIT 5010 GB Rx SENSE H SOURCE H SENSE L SOURCE L RETURN HI POT IR The RETURN terminal is used for IR DCW and ACW tests The SOURCE L H and SENSE L H terminals are for G...

Страница 26: ...e rear panel The IIT 2000 2010 5000 5010 supports line voltages of 100V 120V 220V and 230V Steps 1 Check the line voltage and the fuse in the fuse holder Page 197 The desired line voltage should line up with the arrow on the fuse holder 220 1 0 0 120 230 2 Connect the power cord to the AC voltage input 220 10 0 120 230 3 If the power cord does not have an earth ground ensure the ground terminal is...

Страница 27: ...king 6 Check to make sure the System Self Test passes without errors S y s e m t H a r w a r d e C h C h e c i g n F i r w a r m e C h S Y S T E S M E L F T E S T k e c i g n k e c i g n k After the System Self Test completes the tester will go into VIEW status and be ready to operate I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0...

Страница 28: ... option Follow the instructions below to install the GPIB card WARNING Before installing the optional GPIB card ensure the IIT 2000 2010 5000 5010 is turned off and disconnected from power Steps 1 Remove the screws from the rear panel cover plate 2 Insert the GPIB card into the two slots on either side of the opening Push the card gently until it is fully inserted ...

Страница 29: ... in excess of 5kVAC or 6kVDC Follow all safety precautions warnings and directions given in the following section when using the instrument 1 Only technically qualified personnel should be allowed to operate the safety tester 2 The operating workplace must be fully isolated especially when the instrument is in operation The instrument should be clearly labeled with appropriate warning signage 3 Th...

Страница 30: ...and procedures that must be followed to ensure that the tester is operated in a safe manner WARNING The IIT 2000 2010 5000 5010 generates voltages of up to 5kVAC or 6kVDC Follow all safety precautions warnings and directions given in the following section when using the instrument 1 Never touch the safety tester lead wires terminals probes and other connected equipment when the tester is testing 2...

Страница 31: ... the HIGH VOLTAGE terminal with ground Doing so could charge the chassis to dangerously high voltages 5 Ensure the earth ground of the line voltage is properly grounded 6 Only connect the test leads to the HIGH VOLTAGE SOURCE H SENSE H terminals before the start of a test Keep the test leads disconnected at all other times 7 Always press the STOP button when pausing testing 8 Do not leave the safe...

Страница 32: ...isolated when the instrument is remotely controlled 10 Ensure an adequate discharge time for the DUT When DCW or IR tests are performed the DUT test leads and probes become highly charged The IIT 2000 2010 5000 5010 has discharge circuitry to discharge the DUT after each test The time required for a DUT to discharge depends on the DUT and test voltage Never disconnect the safety tester before a di...

Страница 33: ...free from defects such as cracks or splitting 2 Ensure the safety tester is always connected to an earth ground 3 Test the safety tester operation with a low voltage current output Ensure the safety tester generates a FAIL judgment when the HIGH VOLTAGE and RETURN terminals are shorted using the lowest voltage current as the testing parameters WARNING Do not use high voltages currents when the HIG...

Страница 34: ...Connection 40 ACW DCW IR and GB Manual Testing 42 Choose Recall a Manual Test Number 44 Edit Manual Test Settings 45 Setting the Test Function 46 Setting the Test Voltage or Test Current 47 Setting the Test Frequency 48 Setting the Upper and Lower Limits 49 Setting a Reference Value 52 Setting the Test Time Timer 53 Setting the Ramp Up Time 55 Creating a MANU Test File Name 56 ...

Страница 35: ...ting EDIT Status 71 Running a MANU Test 72 PASS FAIL MANU Test 77 Zeroing of the Test Leads GB only 83 Special MANU Test Mode 000 86 Automatic Tests 92 Choose Recall an Automatic Test 93 Edit Automatic Test Settings 94 Adding a Step to the Automatic Test 95 Creating an AUTO Test File Name 96 Saving and Exiting EDIT Status 98 Automatic Test Page View 99 Running an Automatic Test 102 Automatic Test ...

Страница 36: ...s STOP Press EDIT SAVE Common Utility Settings1 PAGE View AUTO mode only Press PAGE Press ESC Press EDIT SAVE Switch to AUTO mode MANU UTILITY menu MANU mode only 1 Press UTILITY 1 Press EDIT SAVE to save settings or ESC to cancel and return to the previous screen 2 Press the STOP key twice for a FAIL result 3 When in MANU mode selecting MANU number 000 will enter the special manual mode 4 The Swe...

Страница 37: ... 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R A U T O 0 0 1 0 0 2 W VIEW status M M A U N N A _ 2 M A N U 0 0 M A U O T N A _ m A E EDIT status EDIT status is used to edit the manual test or automatic test parameters Pressing the EDIT SAVE key will save any changes Pressing the ESC key will cancel any changes I R G B A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D...

Страница 38: ...us TEST status is active when a MANU test or AUTO test is running Pressing STOP will cancel the MANU test or the remaining steps in an AUTO test I R G B A CW D CW E F R E Q 0 H z 6 0 100k V S E T M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R M M A U N N A _ 2 M A N U 0 0 TEST status T m A 00 33 STOP status STOP status is shown when a manual test or automatic test di...

Страница 39: ...0 1 1 1 2 1 3 1 4 1 5 1 6 1 0 0 m A H I E S 0 T _ AUTO mode AUTO indicates that the tester is in AUTO mode AUTO mode is for creating running a sequence of up to 16 MANU tests I R G B A CW D CW A D D E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M A U O T N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R T 2 A U T O 0 0 3 0 0 AUTO mode m A MANU mode MANU mode is used to create ...

Страница 40: ...D o u b l e A c t i n o L C D B U Z Z N T E I R C T R L F R O N T P A N M O C MO U N T I L I T Y E L F O F I N T E R L C K O O F F MANU Utility Settings The Manu Utility settings are configured for each MANU test separately The settings include ARC MODE PASS HOLD FAIL MODE MAX HOLD and GROUND MODE L I T Y I M A U N U T 2 M A N U 0 0 A R C M D E O O F F G R O U N D MO D E N O P A S S H L D O O F F ...

Страница 41: ...ction Background ACW DCW and IR tests use the HIGH VOLTAGE terminal and RETURN terminal with the GHT 114 test leads ACW DCW IR Connection High Voltage terminal DUT IIT 2010 5010 Return terminal Steps 1 Turn the power off on the safety tester 2 Connect the high voltage test lead red to the HIGH VOLTAGE terminal and screw firmly into place 3 Connect the return test lead white into the RETURN termina...

Страница 42: ...nd SOURCE H L terminals with the GTL 115 test leads GB Connection Source H DUT IIT 2010 5010 Sense L Sense H Source L Steps 1 Turn the power off on the safety tester 2 Connect the Sense H lead to the SENSE H terminal 3 Connect the Sense L lead to the SENSE L terminal 4 Connect the Source H lead to the SOURCE H terminal 5 Connect the Source L lead to the SOURCE L terminal ...

Страница 43: ...IIT 2000 2010 5000 5010 Instruction Manual English 05 08 2015 Version No 001 41 SOURCE H SENSE H SOURCE L SENSE L GB Rx RETURN HI POT IR Sense H lead Sense L lead Source L lead Source H lead ...

Страница 44: ...est can be used as a test step when creating an AUTO test page 92 Choose Recall a Manual Test number from page 44 Edit Manual Test Settings from page 45 Setting the Test Function from page 46 Setting the Test Voltage or Test Current from page 47 Setting the Test Frequency from page 48 Setting the Upper and Lower Limits from page 49 Setting a Reference Value from page 52 Setting the Test Time Timer...

Страница 45: ...Saving and Exiting EDIT Status from page 71 Running a MANU Test from page 72 PASS FAIL MANU Test from page 77 Zeroing of the Test Leads GB only from page 83 Special MANU Test Mode 000 from page 86 Before operating the IIT 2000 2010 5000 5010 please read the safety precautions as outlined in the Set Up chapter on page 24 ...

Страница 46: ... 83 for details on the special mode Steps 1 If the tester is in AUTO mode press and hold the MANU AUTO key for three seconds to switch to MANU mode The tester can only switch between AUTO and MANU mode when in the VIEW status MANU AUTO I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R A U T O 0 0 1 0 0 2 W VIEW status M...

Страница 47: ...r ESC key Edit Manual Test Settings Background To edit any of the manual test settings the tester must be in EDIT status Any settings or parameters that are edited only apply to the currently selected MANU number Steps 1 Press the EDIT SAVE key when in VIEW status to enter the EDIT status This will enter the EDIT status for the chosen test number EDIT SAVE I R G B A CW D CW I L H O I M E T R E F R...

Страница 48: ...function can be set There are four test functions AC Withstand DC Withstand Insulation Resistance and Ground Bond Steps 1 To choose the test function press the ACW DCW IR or GB soft keys A CW D CW I R G B 2 The test function soft key is highlighted I R G B A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A I H E S 0 T 0 0 1 ...

Страница 49: ...to 30A IIT 2010 or 3A to 32A IIT 5010 IIT 5010 includes a 125V test point Steps 1 Press the UP DOWN arrow keys to bring the cursor to the voltage setting I R G B A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 cursor m A 2 Use the scroll wheel to set the voltage level ACW DCW ...

Страница 50: ...ge GBV is calculated as the HI SET limit x Test Current Setting the Test Frequency Background A test frequency of 60Hz or 50Hz can be set regardless of the input line voltage The test frequency setting only applies to ACW and GB tests Steps 1 Press the UP DOWN arrow keys to bring the cursor to the FREQ setting I R G B m A A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0...

Страница 51: ... judged as FAIL When the value exceeds the HI SET setting the test will be judged as FAIL Any measurement between the LO SET and HI SET setting is judged as PASS The LO SET limit cannot be made greater than the HI SET limit Steps 1 Press the HI LO soft key or use the UP DOWN arrow keys to bring the cursor to the HI SET ACW DCW GB setting or the LO SET IR setting I L H O OR I R G B A CW D CW I L H ...

Страница 52: ...021 0mA IIT 5000 5010 0001MΩ 9999MΩ IIT 2010 0 001GΩ 50 00GΩ IIT 5010 000 1mΩ 650 0mΩ 3 Press the HI LO soft key again or press the DOWN arrow key to switch between HI SET and LO SET I L H O OR I R G B A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A 0 L E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 cursor 1 0 0 m A I H E S 0 ...

Страница 53: ...MΩ 9999MΩ IIT 2010 0 001GΩ 50 00GΩ IIT 5010 000 0mΩ 649 9mΩ Note Please note that the resolution of the measured value depends on the resolution of HI SET setting Note The LO SET setting is limited by the HI SET setting The LO SET limit cannot be greater than the HI SET limit When setting the current be aware that a maximum of 200VA can be set for ACW and 50W for DCW IIT 2000 2010 or 500VA and 100...

Страница 54: ...rsor to the REF setting I R G B m A A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A 0 L E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 cursor 2 Use the scroll wheel to set the REF value ACW DCW IR GB 0 000mA HI SET current 0 1mA 0 000mA HI SET current 0 1mA 0000MΩ HI SETΩ 1MΩ 000 0mΩ HI SETΩ 0 1mΩ Note For GB tests a reference...

Страница 55: ...note GB does not have Ramp or discharge times The test time can be set from 0 5 seconds to 999 9 seconds for ACW DCW and GB and 1 0 second to 999 9 seconds for IR with a resolution of 0 1 seconds for all modes The timer can be turned off when in the special MANU test mode when using the ACW or DCW test functions Each test has an initial test time of approximately 150ms and a discharge time except ...

Страница 56: ...or 2 Use the scroll wheel to set the TIMER value ACW DCW IR GB 000 5s 999 9s 000 5s 999 9s 001 0s 999 9s 000 5s 999 9s Note With the ACW test function when the test current is between 30mA and 40mA IIT 2000 2010 or 80mA and 100mA IIT 5010 5000 the ramp time test time cannot exceed 240 seconds At this current level the tester also needs to pause after a test for a time equal to or greater than the ...

Страница 57: ... IIT 2000 2010 or 80mA and 100mA IIT 5010 5000 in ACW mode The discharge time and initial test time cannot be edited Setting the Ramp Up Time Background The Ramp Up time is the total time taken for the tester to reach the test voltage level The Ramp Up time starts after the initial time approximately 150ms with a start voltage of 50 volts The Ramp Up time can be set from 000 1 to 999 9 seconds The...

Страница 58: ... T 2 M A N U 0 0 cursor 2 Use the scroll wheel to set the RAMP value ACW DCW IR 000 1s 999 9s 000 1s 999 9s 000 1s 999 9s Note The discharge time and initial test time cannot be edited Creating a MANU Test File Name Background Each manual test can have a user defined test file name default MANU_NAME up to 10 characters long See the character list below for the allowed characters Character List 0 1...

Страница 59: ...E I R G B m A A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A 0 L E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 cursor 2 Use the scroll wheel to scroll through the available characters 3 Press the Left Right arrow keys to go the next character 4 The MANU test file name is set when the current test setting is saved or when the...

Страница 60: ...ing ACW and DCW testing There are three ARC detection settings OFF ON AND CONTINUE ON AND STOP The ON AND CONTINUE setting will detect arcs over the ARC current level and continue the test the ON AND STOP setting will stop the test when an arc is detected ARC mode settings only apply to ACW and DCW tests Steps 1 Press the UTILITY key on the front panel when the tester is in EDIT status The tester ...

Страница 61: ...o save and exit the MANU Utility and go back to EDIT status EDIT SAVE I R G B m A A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 100k V 0 0 m A R A C 1 0 I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 Note The ESC key can be pressed at any time in the Utility menu to cancel and exit 5 If the ARC MODE was set to either ON AND CON...

Страница 62: ...000 5010 Instruction Manual English 60 05 08 2015 Version No 001 7 Use the scroll wheel to edit the ARC level IIT 2000 2010 ACW DCW 1 000mA 080 0mA 1 000mA 020 0mA IIT 5010 5000 ACW DCW 2 000mA 200 0mA 2 000mA 040 0mA ...

Страница 63: ...A 01 00mA 20 00mA 010 0mA 042 0mA 001 0mA 080 0mA ACW IIT 5010 5000 HI SET Limit ARC Range 0 001mA 1 100mA 2 000mA 01 11mA 11 00mA 02 00mA 20 00mA 011 1mA 110 0mA 002 0mA 200 0mA DCW IIT 2000 2010 HI SET Limit ARC Range 0 001mA 0 999mA 1 000mA 2 000mA 01 00mA 09 99mA 01 00mA 20 00mA 010 0mA 011 0mA 001 0mA 020 0mA DCW IIT 5010 5000 HI SET Limit ARC Range 0 001mA 1 100mA 2 000mA 01 11mA 11 00mA 02 ...

Страница 64: ...g is ignored when running a single MANU test Steps 1 Press the UTILITY key on the front panel when the tester is in EDIT status The display will go from the normal EDIT status to the MANU Utility menu for the current test UTILITY L I T Y I M A U N U T 2 M A N U 0 0 A R C M D E O O F F G R O U N D MO D E N O P A S S H L D O O F F F A I L M D E O P S T O M A X H O L D O F F Note The MANU UTILITY set...

Страница 65: ...o cancel and exit Setting FAIL MODE Background The FAIL MODE settings only apply to the selected test in AUTO tests FAIL MODE has three options CONTINUE HOLD and STOP When FAIL MODE is set to CONTINUE the tester will continue testing after a FAIL judgment When set to HOLD the tester will hold the test on a FAIL judgment and then continue testing after the START key is pressed The STOP mode will co...

Страница 66: ...l EDIT status to the MANU Utility menu for the current test UTILITY L I T Y I M A U N U T 2 M A N U 0 0 A R C M D E O O F F G R O U N D MO D E N O P A S S H L D O O F F F A I L M D E O P S T O M A X H O L D O F F 2 Use the UP DOWN arrow keys to move to the FAIL MODE setting 3 Use the scroll wheel to set FAIL MODE FAIL MODE CONTINUE HOLD STOP 4 Press the EDIT SAVE key to save and exit the MANU Util...

Страница 67: ... the front panel when the tester is in EDIT status The display will go from the normal EDIT status to the MANU Utility menu for the current test UTILITY L I T Y I M A U N U T 2 M A N U 0 0 A R C M D E O O F F G R O U N D MO P A S S H L D O O F F F A I L M D E O P S T O M A X H O L D O F F D E N O Note The MANU UTILITY settings only apply to the selected MANU test 2 Use the UP DOWN arrow keys to mo...

Страница 68: ...on environment This mode measures the potential of the HIGH VOLTAGE terminal with respect to earth ground This means that any stray capacitance resistance that leaks to earth ground will also be measured This is the safest testing mode though potentially not as accurate When GROUND MODE is set to OFF the return terminal is floating with respect to the earth ground This mode is for DUTs that are fl...

Страница 69: ...tion The GROUND MODE is always set to OFF for IR and GB tests GROUND MODE ON DUT grounded High Voltage terminal Return terminal DUT IIT 2000 2010 5000 5010 stray resistance capacitance GROUND MODE ON DUT floating High Voltage terminal DUT IIT 2000 2010 5000 5010 stray resistance capacitance Return terminal ...

Страница 70: ...010 stray resistance capacitance Return terminal IIT 2010 5010 GB testing DUT Source H Source L Sense H Sense L stray resistance capacitance GROUND MODE OFF DUT grounded High Voltage terminal DUT IIT 2000 2010 5000 5010 stray resistance capacitance Return terminal IIT 2010 5010 GB testing DUT Source H Source L Sense H Sense L stray resistance capacitance ...

Страница 71: ... MODE to ON Only set GROUND MODE to OFF when the DUT is floating electrically Steps 1 Press the UTILITY key on the front panel when the tester is in EDIT status The display will go from the normal EDIT status to the MANU Utility menu for the current test UTILITY L I T Y I M A U N U T 2 M A N U 0 0 A R C M D E O O F F G R O U N D MO D E P A S S H L D O O F F F A I L M D E O P S T O M A X H O L D O ...

Страница 72: ... E T R E F R E Q 0 H z 6 0 050k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A 0 L E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 GROUND MODE OFF m A I R G B A CW D CW I L H O I M E T R E F R E Q 0 H z 6 0 050k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A 0 L E S 0 T 0 0 1 S 0 A M P R T 2 M A N U 0 0 GROUND MODE ON m A Note The ESC key can be pressed ...

Страница 73: ...nnot be used for AUTO tests See page 83 for details Steps 1 When in EDIT status press the EDIT SAVE key to save the current test This will enter the VIEW status for the chosen test number EDIT SAVE I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W 2 M A N U 0 0 I D E T m A 2 The Status changes from EDI...

Страница 74: ...on the screen See page 198 for a comprehensive list of the all the setting errors The INTERLOCK function is ON and the Interlock key is not inserted in the signal I O port page 117 The STOP signal has been received remotely If Double Action is ON ensure the START button is pressed immediately after the STOP button 0 5s Note When a test is running the voltage output cannot be changed unless the tes...

Страница 75: ...ter into the READY status STOP I R G B m A A CW D CW E F R E Q 0 H z 6 0 100k V A E R M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R D M M A U N N A _ 2 M A N U 0 0 READY status Y 3 The READY indicator will be lit blue when in the READY status READY 4 Press the START button when the tester is in the READY status The manual test starts automatically and the tester goe...

Страница 76: ...stopped I R G B A CW D CW E F R E Q 0 H z 6 0 100k V S E T M T I E 0 0 3 2 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R M M A U N N A _ 2 M A N U 0 0 T remaining TIMER time 00 37m A remaining RAMP time ACW Example I R G B A CW D CW E F R E Q 0 H z 6 0 100k V S E T M T I E 0 0 3 2 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R M M A U N N A _ 2 M A N U 0 0 T Test Vol...

Страница 77: ...3 T M M A U N N A _ 2 M A N U 0 0 T Test Current Measured Resistance 200 7mΩ Stop the Test 1 To stop the test at any time when it is running press the STOP button The test will stop immediately When the STOP button is pressed a judgment is not made on the test All panel keys except the STOP button are locked when the tester is in STOP status STOP I R G B A CW D CW E F R E Q 0 H z 6 0 100k V O T S ...

Страница 78: ...er is in the READY status The tester will revert to the VIEW status for the current test MANU AUTO I R G B m A A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M M A U N N A _ 2 M A N U 0 0 Note Do not touch any terminals test leads or any other connections when the test is on ...

Страница 79: ... will be judged FAIL when Either the HI SET or LO SET limit has been tripped during the test time A protection setting has been tripped during the test time See page 198 for a list of error messages PASS Judgment 1 When the test is judged as PASS PASS will be displayed the buzzer will sound and the PASS indicator will be lit green PASS I R G B A CW D CW E F R E Q 0 H z 6 0 100k V S A P M T I E 0 0...

Страница 80: ... test START Note The buzzer will only sound if the Pass Sound is set to ON See page 113 for details The START button is disabled when the buzzer is beeping PASS Timing Diagrams The timing diagrams below show the ACW DCW IR and GB timing for the START status TEST status and PASS judgment ACW PASS Timing Discharge time Output V time Initial time START TEST PASS RAMP TEST TIME Approximately 150ms ...

Страница 81: ...ischarge time Output V time Initial time START TEST PASS RAMP TEST TIME Approximately 150ms IR PASS Timing Discharge time Output V time Initial time START TEST PASS RAMP TEST TIME Approximately 150ms GB PASS Timing Output I time START TEST PASS 0 5s Initial time Total Test Time Approximately 150ms ...

Страница 82: ...1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R 2 M A U N 0 0 _ 2 M A N U 0 0 L m A 01 37 2 The FAIL judgment will be held on the display until the STOP button is pressed Pressing the STOP button twice will return the tester to the READY status STOP 3 The READY indicator will be lit blue in the READY status READY I R G B m A A CW D CW E F R E Q 0 H z 6 0 100k V A E R M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 ...

Страница 83: ... START status TEST status and FAIL judgment ACW FAIL Timing START TEST FAIL Discharge time Output V time Initial time RAMP TEST TIME Approximately 150ms DCW FAIL Timing FAIL Discharge time Output V time Initial time RAMP START TEST TEST TIME Approximately 150ms IR FAIL Timing START TEST FAIL Discharge time Output V time Initial time RAMP TEST TIME Approximately 150ms ...

Страница 84: ...IIT 2000 2010 5000 5010 Instruction Manual English 82 05 08 2015 Version No 001 GB FAIL Timing Output I time START TEST FAIL 0 5s Initial time Total Test Time Approximately 150ms ...

Страница 85: ... reference is automatically set to the measured resistance of the test leads This function is only available for GB testing Steps 1 Ensure the tester is in VIEW status for the current GB test Save the current test if necessary Page 71 I R G B A CW D CW E F R E Q 0 H z 6 E I V G B V V M T I E 0 0 3 2 S R 0 mΩ E R F 0 0 0 0 0 0 mΩ I H E S 3 T M M A U N N A _ 2 M A N U 0 0 W mΩ 03 01A 2 Short the pos...

Страница 86: ...5 Press the START button to perform the zero check The tester will go into the ZERO status START 03 01A Z E R 0 I R G B A CW D CW E F R E Q 0 H z 6 R E Z G B V 0 6 2 2 V M T I E 0 0 3 2 S R 0 mΩ E R F 0 0 0 0 0 0 mΩ I H E S 3 T M M A U N N A _ 2 M A N U 0 0 O Measured Resistance 200 7mΩ ZERO status 6 When the zero check has finished the tester will judge the test as either PASS or FAIL If the test...

Страница 87: ... test will fail and an I SET error will appear on the screen Stop the test and re check the connection again and try again Z E R 0 I R G B A CW D CW E F R E Q 0 H z 6 I A F G B V 5 6 2 2 V M T I E 0 0 3 2 S R 0 mΩ E R F 0 0 0 0 0 0 mΩ I H E S 3 T M M A U N N A _ 2 M A N U 0 0 L I SET error message FAIL status I S E T R 0 If an R 0 error message appear on the screen stop the test and perform the ze...

Страница 88: ...nctions ACW DCW IR and GB This means a different ACW DCW IR and GB test setup can be saved for MANU number 000 Sweep Function Overview The IIT 5000 and 5010 have access to the sweep mode function The sweep function creates a graph of one of the ACW DCW IR or GB tests in the special manual mode The graph will plot the output voltage current or resistance versus time After the test has been complete...

Страница 89: ...ltage Current The test items that are plotted on the sweep graph depend on the type of test that is performed TEST Graph Test Items ACW Test voltage measured current V I DCW Test voltage measured current V I IR Test voltage measured resistance V R GB Test current measured resistance I R Steps 1 Choose MANU number 000 to enter the special test mode Page 44 ...

Страница 90: ...mple ACW 3 Set all the necessary parameters for a test and save Note A different test setup can be saved for each test function ACW DCW IR and GB Pages 45 71 I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V S T A t 0 0 0 0 1 S M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M M A U N N A _ 0 M A N U 0 0 m A E SW E S T A t P IIT 5010 2010 shown Note The TIMER setti...

Страница 91: ... 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M M A U N N A _ 0 M A N U 0 0 m A E SW E S T A t P Start time 2 Press the EDIT SAVE key to save the Start time EDIT SAVE Running the Test 1 In special test mode 000 tests are started and stopped in the same way as for the normal manual test mode See page 72 for details Page 72 2 If required the scroll wheel can be used to set the voltage level in real t...

Страница 92: ...lable for the IIT 5000 5010 Steps 1 When the test has finished press the SWEEP key to view the results of the sweep in a graph WE E S P Graph Test Items TEST Primary Secondary ACW Test voltage measured current DCW Test voltage measured current IR Test voltage measured resistance GB Test current measured resistance DCW Example D CW Primary item Secondary item I m A V k V 0 2 0 0 0 0 5 1 T S 0 0 0 4...

Страница 93: ...e are shown on the left hand side Remove Lines from the Graph 1 Pressing the F5 key will toggle the primary test item on off 2 Pressing the F6 key will toggle the secondary test item on off D CW Test Voltage I m A V k V 0 2 0 0 0 0 5 1 T S 0 0 0 4 5 V I Test Time Toggle voltage line Toggle current line Example DCW test Exit the Results Graph To exit the graph press the ESC key You will be returned...

Страница 94: ...st step when creating an AUTO test Choose Recall an Automatic Test from page 93 Edit Automatic Test Settings from page 94 Adding a Step to the Automatic Test from page 95 Creating an AUTO Test File Name from page 96 Saving and Exiting EDIT Status from page 98 Automatic Test Page View from page 99 Running an Automatic Test from page 102 Automatic Test Results from page 107 Before operating the IIT ...

Страница 95: ... will put the tester into Auto mode The tester can only switch between AUTO and MANU mode when in the VIEW status MANU AUTO I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R M A N U 0 0 1 W M A U O T N A _ 0 A U T O 0 0 1 1 0 VIEW status M M A U N N A _ E m A Note If the chosen automatic test has not yet been setup then...

Страница 96: ... 0 1 1 0 AUTO number m A Note The AUTO number can only be chosen in VIEW status If in the EDIT status switch to the VIEW status by pressing the EDIT SAVE or ESC key Edit Automatic Test Settings Background To edit an automatic test the tester must be in EDIT status Any settings or parameters that are edited only apply to the currently selected AUTO number Steps 1 Press the EDIT SAVE key when in VIE...

Страница 97: ...g the EDIT SAVE key again will save the settings or pressing the ESC will cancel the settings for the current AUTO test and return back to VIEW status Adding a Step to the Automatic Test Background Up to 16 MANU tests steps can be added to an automatic AUTO test Each step is added in a sequential order Steps 1 Press the DOWN arrow keys to bring the cursor to the MANU number I R G B cursor MANU num...

Страница 98: ...dded to an AUTO test FULL will be shown on the display when you attempt to add another step to the AUTO test I R G B m A A CW D CW A D D E F R E Q 0 H z 6 0 100k V L U F M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M M A U N N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R L 1 A U T O 0 0 1 0 0 Note The test order can be edited in the Page View menu after the AUTO test is saved See page 99 for details Creat...

Страница 99: ...initially set as AUTO_NAME I R G B m A A CW D CW A D D E F R E Q 0 H z 6 0 100k V I D E M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M A U O T N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R T 1 A U T O 0 0 1 0 0 cursor AUTO test file name 2 Use the scroll wheel to scroll through the available characters 3 Press the LEFT RIGHT arrow keys to go to the next character 4 The AUTO test file name is set when the...

Страница 100: ... EDIT SAVE key to save the automatic test After the test is saved the tester will revert back to VIEW status EDIT SAVE I R G B m A A CW D CW A D D E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 M A U O T N A _ 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W 1 A U T O 0 0 1 0 0 I D E T 2 The status changes from EDIT to VIEW Note Pressing the EDIT SAVE key again will return the tes...

Страница 101: ...eps 1 Ensure the tester has had an automatic test saved and the tester is in AUTO mode VIEW status Page 92 I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W VIEW status M A U O T N A _ 0 A U T O 0 0 1 1 0 AUTO mode m A 2 Press the PAGE key to bring up the Page view of the AUTO test All the test steps are shown on the ...

Страница 102: ... be edited Steps can be deleted skipped moved or swapped Moving a Step 1 Use the UP DOWN and LEFT RIGHT arrow keys to move the cursor to the test step you wish to move 2 Press the MOVE soft key O V E M 3 Use the UP DOWN and LEFT RIGHT arrow keys to move the cursor to the destination step 4 Press the MOVE soft key again The manual test will be moved to the destination step The remaining steps will ...

Страница 103: ... V E M WA P S K I P S D E L E M A U O T N A _ 0 A U T O 0 0 1 0 1 M A N U _ N M E A 0 1 0 1 0 0 2 0 0 1 0 3 0 0 3 0 4 0 0 4 0 5 0 0 7 0 6 0 0 3 0 7 0 3 8 0 8 0 0 5 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 Skip a Test Step 1 Use the UP DOWN and LEFT RIGHT arrow keys to move the cursor to the test step you wish to skip 2 Press the SKIP soft key K I P S 3 The step will have an asterisk beside the MANU number ...

Страница 104: ... DEL soft key D E L 3 The step will be deleted Save Changes and Exit To save the changes made in Page View press the EDIT SAVE key You will be returned back to AUTO mode VIEW status EDIT SAVE Cancel and Exit Page View To cancel any changes and to exit the Page View press the ESC key You will be returned back to AUTO mode VIEW status ESC Running an Automatic Test Background An automatic test can be...

Страница 105: ... is ON ensure the START button is pressed immediately after the STOP button 0 5s Warning Do not touch any terminals test leads or the DUT when a test is running Steps 1 Ensure the tester is in VIEW status Save the automatic test if necessary Page 93 I R G B A CW D CW E F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M A U O T N A _ 0 A...

Страница 106: ... is in the READY status The AUTO test starts automatically and the display changes to TEST status START 5 The TEST indicator will be lit orange when in the TEST status TEST 6 Each test will start by showing the remaining ramp up time followed by the remaining test time Each test will be tested in sequence until the last test has finished or the test is stopped I R G B A CW D CW F R E Q 0 H z 6 0 1...

Страница 107: ...H z 6 0 100k V L O H M T I E 0 0 3 2 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R D 00 37m A E M A U O T N A _ 1 A U T O 0 0 1 0 0 HOLD status 2 The PASS or FAIL indicator will also be lit The buzzer will NOT sound PASS FAIL 3 To continue to the next test after HOLD is displayed on screen press the START button START 4 To stop the test when HOLD is displayed on screen press the ST...

Страница 108: ...aborted All panel keys except the STOP and START buttons are locked when the tester has been stopped All the results up until when the AUTO test was stopped are shown on screen See page 107 for more details on automatic test results STOP E M A U O T N A _ A U T O 0 0 1 0 1 F A I 0 2 P A S 0 3 S T O 0 4 0 5 0 6 0 7 0 8 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 L S P Example of an automatic test that has been...

Страница 109: ...lowed to run to completion the AUTO test is not stopped or a protection setting is not tripped then the tester will judge each step as either PASS or FAIL This is shown as a table after the automatic test has finished running If the test has been stopped then any remaining tests will not be run and thus the AUTO test will not finish running Overview P A S P A S S S P A S S K I S P E M A U O T N A ...

Страница 110: ...as neither PASS nor FAIL No step can contain an ERROR or ILOCK message E R R P A S O S P A S S K I S P E M A U O T N A _ 1 A U T O 0 0 1 0 0 0 1 0 2 0 3 0 5 0 6 0 7 0 8 S T O 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 P I L O F A I C L ILOCK message ERROR message R K ERROR Indicates that V I or R is not correct This usually occurs if the testing leads are not properly connected ILOCK Indicates that the inter...

Страница 111: ...zzer will sound FAIL P A S P A S S S P A S F A I S L E M A U O T N A _ A U T O 0 0 1 0 1 0 2 0 3 0 4 P A S 0 5 0 6 0 7 0 8 P A S 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 S S P A S F A I S L Note The Fail Sound setting must to set to ON for the buzzer to sound page 114 View Results 1 When the PASS or FAIL overview table is shown on the screen turn the scroll wheel right to scroll through each test step I R ...

Страница 112: ...ADY indicator will be lit blue in the READY status READY I R G B A CW D CW E F R E Q 0 H z 6 0 100k V A E R M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R Y READY status M A U O T N A _ 0 A U T O 0 0 1 1 0 D m A Exit Testing To exit testing press the MANU AUTO key when the tester is in the READY status The tester will revert to the VIEW status MANU AUTO I R G B A CW ...

Страница 113: ...2000 2010 5000 5010 Instruction Manual English 05 08 2015 Version No 001 111 PASS Timing Diagram Output V time START TEST PASS step 1 step 16 FAIL Timing Diagram Output V time START TEST FAIL step 1 step 16 ...

Страница 114: ...age 112 Buzzer Settings from page 113 Interface Settings from page 115 Control settings from page 117 LCD Settings Description The LCD settings include contrast and brightness controls Steps 1 Ensure the tester is in VIEW status Save the current test if necessary Page 71 I R G B E A CW D CW F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R...

Страница 115: ... I R C T R L s B R I G H T s t 5 M O C MO U N T I L I T Y 4 Use the UP DOWN arrow keys to choose a menu item LCD Contrast LCD Brightness 5 Use the scroll wheel to select a parameter for the chosen menu item LCD Contrast LCD Brightness 1 low 8 high BRIGHT DARK 6 Press EDIT SAVE to save the settings and exit to VIEW status EDIT SAVE Note The ESC key can be pressed at any time to cancel and exit back...

Страница 116: ... status Save the current test if necessary Page 71 I R G B E A CW D CW F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M M A U N N A _ 0 M A N U 0 0 m A VIEW status 2 Press the UTILITY key UTILITY 3 Press the BUZZ soft key to bring up the Buzzer Common Utility menu B U Z Z L C D B U Z Z N T E I R C T R L M O C MO U N T I L I T Y P a s ...

Страница 117: ...ote When in automatic tests the Pass Sound and Fail Sound settings only apply to the overall PASS FAIL of the overall automatic test not each test step that make up the automatic tests Note The ESC key can be pressed at any time to cancel and exit back to VIEW status Interface Settings Description The interface settings choose the remote interface configuration USB RS232 and GPIB optional can be s...

Страница 118: ...Press the UTILITY key UTILITY 3 Press the INTER soft key to bring up the Interface Common Utility menu N T E I R I n t L C D B U Z Z N T E I R C T R L M O C MO U N T I L I T Y B u d a 1 1 5 2 0 0 R S 2 3 2 f a c e e r 4 Use the scroll wheel to select USB RS232 or GPIB 5 For RS232 or GPIB use the UP DOWN arrow keys to choose Baud or Address 6 Use the scroll wheel to select the baud rate or GPIB add...

Страница 119: ...ontrol settings include Start Control Double Action Key Lock and Interlock Start Control is used to determine how a test is started Tests can be started via the front panel START STOP buttons from a remote controller or via the SIGNAL I O port The Double Action function is a safety feature used to prevent accidentally starting a test Normally to start a test the START button is pressed when the te...

Страница 120: ...used for this purpose See page 127 for details Steps 1 Ensure the tester is in VIEW status Save the current test if necessary Page 71 I R G B E A CW D CW F R E Q 0 H z 6 0 100k V E I V M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R W M M A U N N A _ 0 M A N U 0 0 m A VIEW status 2 Press the UTILITY key UTILITY 3 Press the CTRL soft key to bring up the Control Common ...

Страница 121: ...NT PANEL REMOTE CONNECT SIGNAL IO ON OFF ON OFF ON OFF 6 Press EDIT SAVE to save the settings and exit to VIEW status EDIT SAVE Note The Double Action setting is ignored when the IIT 2000 2010 5000 5010 is being controlled remotely using the USB RS232 or GPIB interfaces Note If a test is started with INTERLOCK ON but the interlock signal I O pins are not shorted either with the included interlock ...

Страница 122: ...ish 120 05 08 2015 Version No 001 I R G B m A A CW D CW E F R E Q 0 H z 6 0 100k V A E R M T I E 0 0 1 0 S R 0 0 m A E R F 0 0 1 0 0 m A I H E S 0 T 0 0 1 S 0 A M P R D M M A U N N A _ 2 M A N U 0 0 Y I N T E R L O C K O P E N Interlock open message ...

Страница 123: ...XTERNAL CONTROL The External Control chapter covers the REMOTE terminal and the SIGNAL I O port External Control Overview 122 Remote Terminal Overview 122 Remote Controller Operation 123 SIGNAL I O Overview 124 Using the SIGNAL I O to Start Stop Tests 127 Using the Interlock Key 128 ...

Страница 124: ...w The REMOTE terminal connector is a standard 5 pin DIN terminal suitable for a remote controller WARNING Keep any cables that are connected to the REMOTE terminal away from the HIGH VOLTAGE and RETURN terminals Pin Assignment and Connection REMOTE 1 4 2 5 3 RMT_START RMT_STOP Pin Pin name Description 1 RMT_STOP Remote Stop signal 2 COM Common line 3 Not used 4 RMT_START Remote Start signal 5 Not ...

Страница 125: ...TOP button To use the REMOTE terminal the IIT 2000 2010 5000 5010 must first be configured to accept a remote controller Operating a remote controller is the same as operating the START and STOP buttons on the front panel Steps 1 Insert the lead of remote controller into the REMOTE terminal REMOTE 2 Configure the Start Ctrl option to REMOTE CONNECT in the Common Utility menu Page 117 3 The tester ...

Страница 126: ...rview Overview The SIGNAL I O port can be used to remotely start stop tests and monitor the test status of the instrument The SIGNAL I O port is also used for the interlock function page 117 The SIGNAL I O port uses a DB 9 pin female connector Pin Assignment 6 9 7 8 1 2 3 4 5 Pin name Pin Description INTERLOCK1 1 When INTERLOCK is ON a test is only allowed to start when both INTERLOCK pins are sho...

Страница 127: ...TERLOCK2 PIN 1 INTERLOCK1 Input Connection PIN 5 INPUT_STOP PIN 3 INPUT_COM PIN 4 INPUT_START Output Connection PIN 6 OUTPUT_TEST PIN 9 OUTPUT_COM PIN 8 OUTPUT_PASS PIN 7 OUTPUT_FAIL Signal Properties Input Signals High level input voltage 5V 32V Low level input voltage 0V 1V Low level input current Maximum of 5mA Input period Minimum of 1ms Output Signals Output Type Relay form A Output Rated Vol...

Страница 128: ...IIT 2000 2010 5000 5010 Instruction Manual English 126 05 08 2015 Version No 001 Input Stop and Input Start Timing INPUT_STOP INPUT_START ...

Страница 129: ... 117 2 Connect the Input Output signals to the SIGNAL I O port 3 To start the testing short the INPUT_STOP and INPUT_COM line for a minimum of 1ms to put the tester into READY status 4 To start the testing short the INPUT_START and INPUT_COM lines for a minimum of 1ms 5 To stop the testing temporarily short the INPUT_STOP and INPUT_COM line again NOTE Even if the IIT 2000 2010 5000 5010 is configu...

Страница 130: ...1 and INTERLOCK2 pins on the signal I O port See page 124 for the Signal I O pin assignment Panel operation 1 Insert the Interlock key into the SIGNAL I O port on the rear panel 2 Set the INTERLOCK option to ON in the Common Utility Page 117 Note With INTERLOCK set to ON the tester can now only start a test when the Interlock key is connected Do not remove the interlock after starting a test It mu...

Страница 131: ...05 08 2015 Version No 001 129 REMOTE CONTROL This chapter describes basic configuration of IEEE488 2 based remote control The remote interface supports USB RS232 and GPIB Interface Configuration 130 Command Syntax 136 Command List 139 Error Messages 194 ...

Страница 132: ...ions device class VCP Virtual Com Port Panel operation 1 Connect the USB cable to the rear panel USB A port 2 Set the interface to USB from the Common Utility menu Page 115 Note When USB is used for remote control an RS232 port is simulated Check the Windows Device Manager for the baud rate and other RS232 settings Check the RS232 configuration below for more details Note the baud rate is fixed to...

Страница 133: ...200 38400 57600 115200 Parity None Data bits 8 Stop bit 1 Flow control None Pin Assignment 1 2 3 4 5 6 7 8 9 1 No connection 2 RxD Receive Data 3 TxD Transmit Data 4 No connection 5 GND 6 9 No connection Connection PC Tester DB9 Pin Signal Signal DB9Pin 2 RxD TxD 3 3 TxD RxD 2 5 GND GND 5 Panel operation 1 Connect the Null modem cable to the rear panel RS232 port RS232 ...

Страница 134: ... Set the interface to RS232 from the Common Utility menu Page 115 GPIB Remote Interface GPIB Configuration Address 0 30 Panel operation 1 Connect the GPIB cable to the rear panel GPIB port GPIB 2 Set the interface to GPIB and set the GPIB address from the Common Utility menu Page 115 ...

Страница 135: ...WinXP Control panel System Hardware tab Run this query command via the terminal after the instrument has been configured for USB or RS232 remote control page 130 130 idn This should return the Model number Serial number and Firmware version in the following format IIT 2000 XXXXXXXXXXXX V1 00 Model number IIT 2000 Serial number 12 character serial number Firmware version V1 00 j can be used as the ...

Страница 136: ...0 0 _ 2 M A N U 0 0 m A Return to Panel Control Background When the instrument is remotely controlled all panel keys except the STOP button are disabled Steps 1 When RMT is on the display press the STOP button The panel goes to the READY status STOP 2 From the READY status the tester can go into one of two states TEST or VIEW To put the tester into VIEW status press the MANU AUTO key To put the te...

Страница 137: ...00 2010 5000 5010 Instruction Manual English 05 08 2015 Version No 001 135 automatic test see pages 72 and 102 respectively START Note To put the tester back to RMT simply issue another remote control command ...

Страница 138: ...mand tree is a node Each keyword in an SCPI command represents each node in the command tree Each keyword node of an SCPI command is separated by a colon For example the diagram below shows an SCPI sub structure and a command example MANU ACW CLOSet VOLTage CHISet MANU ACW VOLTage Command types There are a number of different instrument commands and queries A command sends instructions or data to ...

Страница 139: ...ms Commands and queries have two different forms long and short The command syntax is written with the short form of the command in capitals and the remainder long form in lower case The commands can be written in capitals or lower case just so long as the short or long forms are complete An incomplete command will not be recognized Below are examples of correctly written commands Long form SYSTem...

Страница 140: ...100 1 Command header 2 Space 3 Parameter Parameters Type Description Example Boolean Boolean logic 0 1 NR1 integers 0 1 2 3 NR2 decimal numbers 0 1 3 14 8 5 NR3 floating point 4 5e 1 8 25e 1 NRf any of NR1 2 3 1 1 5 4 5e 1 string ASCII text string TEST_NAME Message Terminator CR LF Carriage Return Line feed code ...

Страница 141: ...tness 144 SYSTem BUZZer PSOUND 144 SYSTem BUZZer FSOUND 145 SYSTem BUZZer PTIMe 145 SYSTem BUZZer FTIMe 146 SYSTem ERRor 146 SYSTem GPIB VERSion 148 Function Commands FUNCtion TEST 149 MEASure x 150 MAIN FUNCtion 151 Manual Commands MANU STEP 154 MANU NAME 155 MANU INITial 155 MANU RTIMe 156 MANU EDIT MODE 157 MANU ACW VOLTage 157 MANU ACW CHISet 158 ...

Страница 142: ... MANU DCW CHISet 163 MANU DCW CLOSet 164 MANU DCW TTIMe 165 MANU DCW REF 165 MANU DCW ARCCurrent 166 MANU IR VOLTage 167 MANU IR RHISet 167 MANU IR RLOSet 168 MANU IR TTIMe 169 MANU IR REF 170 MANU GB CURRent 171 MANU GB RHISet 171 MANU GB RLOSet 172 MANU GB TTIMe 172 MANU GB FREQuency 173 MANU GB REF 173 MANU GB ZEROCHECK 174 MANU UTILity ARCMode 174 MANU UTILity PASShold 175 MANU UTILity FAILmod...

Страница 143: ...T SHOW 177 Sweep Commands SWEEP DATA STATus 179 SWEEP X DATA SHOW 180 SWEEP GRAPh SHOW 181 SWEEP GRAPh LINE 182 SWEEP STARt TIME 183 Auto Commands AUTO STEP 184 AUTO x PAGE SHOW 185 AUTO PAGE MOVE 185 AUTO PAGE SWAP 186 AUTO PAGE SKIP 187 AUTO PAGE DEL 187 AUTO NAME 188 AUTO EDIT ADD 188 TESTok RETurn 189 SRE 190 Common Commands CLS 191 IDN 191 ...

Страница 144: ...IIT 2000 2010 5000 5010 Instruction Manual English 142 05 08 2015 Version No 001 Remote Commands RMTOFF 192 Special Functions Inter Lock Key Open 192 ...

Страница 145: ...4 SYSTem BUZZer FSOUND 145 SYSTem BUZZer PTIMe 145 SYSTem BUZZer FTIMe 146 SYSTem ERRor 146 SYSTem GPIB VERSion 148 SYSTem LCD CONTrast Set Query Description Sets the contrast of the LCD display from 1 low to 8 bright Syntax Query Syntax SYSTem LCD CONTrast NR1 SYSTem LCD CONTrast Parameter Return parameter NR1 1 8 Example SYST LCD CONT 5 Sets the display contrast to 5 ...

Страница 146: ...1 SYSTem LCD BRIGhtness Parameter Return parameter NR1 1 dark 2 bright Example SYST LCD BRIG 2 Sets the display brightness to bright SYSTem BUZZer PSOUND Set Query Description Turns the buzzer sound on or off for a PASS judgment Syntax Query Syntax SYSTem BUZZer PSOUND ON OFF SYSTem BUZZer PSOUND Parameter Return parameter ON PASS Sound on OFF PASS Sound off Example SYST BUZZ PSOUND ON Turns the b...

Страница 147: ... OFF SYSTem BUZZer FSOUND Parameter Return parameter ON FAIL Sound on OFF FAIL Sound off Example SYST BUZZ FSOUND ON Turns the buzzer sound on for FAIL judgments SYSTem BUZZer PTIMe Set Query Description Sets the PASS sound duration in seconds Syntax Query Syntax SYSTem BUZZer PTIMe NR2 SYSTem BUZZer PTIMe Parameter Return parameter NR2 0 2 999 9 Example SYST BUZZ PTIM 1 Sets the buzzer to 1 secon...

Страница 148: ...UZZer FTIMe NR2 SYSTem BUZZer FTIMe Parameter Return parameter NR2 0 2 999 9 Example SYST BUZZ FTIM 1 Sets the buzzer to 1 second for a FAIL judgment SYSTem ERRor Query Description Returns any errors in the output buffer See the error code table below for details Query Syntax SYSTem ERRor Return p a r a m e string Returns an error string that includes an error code and an error description ...

Страница 149: ...Query Error 24 Mode Error 25 Time Error 26 DC Over 50W IIT 2000 2010 DC Over 100W IIT 5010 5000 27 GBV 5 4V 30 Voltage Setting Error 31 Current Setting Error 32 Current HI SET Error 33 Current LO SET Error 34 Resistance HI SET Error 35 Resistance LO SET Error 36 REF Setting Error 37 Frequency Setting Error 38 ARC Setting Error 39 RAMP Time Setting Error 40 TEST Time Setting Error 45 Buffer Error ...

Страница 150: ...0 gets SWEEP data Example SYST ERR 0 No Error Returns 0 No Error as the error message SYSTem GPIB VERSion Query Description Queries the GPIB version Query Syntax SYSTem GPIB VERSion Return parameter string Returns The GPIB version as a string GPIB V1 00 or No GPIB connected if there is not a GPIB device configured connected Query Example SYST GPIB VERS GPIB V1 00 Returns the GPIB version ...

Страница 151: ...st output on or off When HOLD is displayed on the screen during AUTO tests use the FUNCtion TEST command to move on to the next step Setting the FUNCtion TEST command to OFF at the end of a test will also temporarily turn the PASS FAIL buzzer sound off Syntax Query Syntax FUNCtion TEST ON OFF FUNCtion TEST Parameter ON Turns the test on OFF Turns the test off Return parameter TEST ON Test is on TE...

Страница 152: ...ers results of the selected step 1 16 of the AUTO test Return parameters function judgment status test voltage test current resistance test time time of completed test or ramp time elapsed time of test that has not been completed Query Syntax MEASure x Parameter MANU mode No parameter needed for MANU mode Parameter AUTO mode x NR1 1 16 Step number Return parameter string Returns the test status of...

Страница 153: ...mple in MANU mode MEAS ACW FAIL 0 024kV 0 013 mA R 000 1S Returns the test result of the current manual test Example in MANU mode MEAS IR TEST 0 250kV Mohm T 000 2S Returns as the test result when the reading is invalid Example in AUTO mode MEAS10 IR FAIL 0 250kV 999M ohm T 010 3S Returns step 10 of the current automatic result MAIN FUNCtion Set Query Description Changes the mode between AUTO and ...

Страница 154: ... 5000 5010 Instruction Manual English 152 05 08 2015 Version No 001 Parameter Return parameter MANU Puts the tester mode to MANU AUTO Puts the tester mode to AUTO Example MAIN FUNC MANU Sets the tester to MANU mode ...

Страница 155: ...MANU EDIT MODE 157 MANU ACW VOLTage 157 MANU ACW CHISet 158 MANU ACW CLOSet 158 MANU ACW TTIMe 159 MANU ACW FREQuency 160 MANU ACW REF 161 MANU ACW ARCCurrent 162 MANU DCW VOLTage 162 MANU DCW CHISet 163 MANU DCW CLOSet 164 MANU DCW TTIMe 165 MANU DCW REF 165 MANU DCW ARCCurrent 166 MANU IR VOLTage 167 MANU IR RHISet 167 MANU IR RLOSet 168 MANU IR TTIMe 169 ...

Страница 156: ...y 173 MANU GB REF 173 MANU GB ZEROCHECK 174 MANU UTILity ARCMode 174 MANU UTILity PASShold 175 MANU UTILity FAILmode 175 MANU UTILity MAXHold 176 MANU UTILity GROUNDMODE 177 MANU x EDIT SHOW 177 MANU STEP Set Query Description Sets the MANU test number Syntax Query Syntax MANU STEP NR1 MANU STEP Parameter Return parameter NR1 0 100 Example MANU STEP 100 Sets the manual test number to 100 ...

Страница 157: ... set the MANU test name Syntax Query Syntax MANU NAME string MANU NAME Parameter Return parameter string 10 character string first character must be a letter Example MANU NAME test1 Sets the manual test name to test1 MANU INITial Set Query Description Loads the initial default settings for the selected MANU test number The initial settings that are loaded depend on the test function ACW DCW IR or ...

Страница 158: ... Loads the initial settings for the selected MANU number MANU RTIMe Set Query Description Sets or returns the Ramp Time for the test in seconds Note A TIME ERR will result if the Ramp Time Test Time is 240 seconds when the HI SET limit is over 30mA IIT 2000 2010 or over 80mA IIT 5010 5000 This applies to the ACW function only Syntax Query Syntax MANU RTIMe NR2 MANU RTIMe Parameter Return parameter...

Страница 159: ...B MANU EDIT MODE Parameter Return parameter ACW AC Withstand mode DCW DC Withstand mode IR Insulation Resistance mode GB Ground Bond mode Example MANU EDIT MODE ACW Sets the mode to ACW MANU ACW VOLTage Set Query Description Sets or returns the ACW voltage in kV The test must first be in ACW mode before this command can be used Syntax Query Syntax MANU ACW VOLTage NR2 MANU ACW VOLTage Parameter Re...

Страница 160: ...U ACW CHISet NR2 MANU ACW CHISet Parameter Return parameter NR2 0 001 042 0 IIT 2000 2010 0 001 110 0 IIT 5010 5000 Example MANU ACW CHIS 10 0 Sets the ACW HI SET current to 10 mA MANU ACW CLOSet Set Query Description Sets or returns the ACW LO SET current value in milliamps The LO SET value must be less than the HI SET value The test must first be in ACW mode before this command can be used The L...

Страница 161: ...ts are outside the range of HI SET LO SET value2 will not produce an error but will return 0 05 not 0 053 Syntax Query Syntax MANU ACW CLOSet NR2 MANU ACW CLOSet Parameter Return parameter NR2 0 000 041 9 IIT 2000 2010 0 000 109 9 IIT 5010 5000 Example MANU ACW CLOS 20 0 Sets the ACW LO SET current to 20 mA MANU ACW TTIMe Set Query Description Sets or returns the ACW test time in seconds The test ...

Страница 162: ...Me NR2 OFF MANU ACW TTIMe Parameter NR2 OFF 0 5 999 9 seconds TIMER OFF special MANU mode Return parameter NR2 TIME OFF 0 5 999 9 seconds TIMER is OFF special MANU mode Example MANU ACW TTIM 1 Sets the ACW test time to 1 second MANU ACW FREQuency Set Query Description Sets or returns the ACW test frequency in Hz The test must first be in ACW mode before this command can be used Syntax Query Syntax...

Страница 163: ...erence value in mA The test must first be in ACW mode before this command can be used The ACW reference value must be less than the HI SET value The ACW reference value must use the same range as the HI SET value Syntax Query Syntax MANU ACW REF NR2 MANU ACW REF Parameter Return parameter NR2 0 000 041 9 IIT 2000 2010 0 000 109 9 IIT 5010 5000 Example MANU ACW REF 0 01 Sets the ACW reference to 0 ...

Страница 164: ... current is limited to 2X the HI SET value Syntax Query Syntax MANU ACW ARCCurrent NR2 MANU ACW ARCCurrent Parameter Return parameter NR2 1 000 080 0 IIT 2000 2010 2 000 200 0 IIT 5010 5000 Example MANU ACW ARCC 0 04 Sets the ACW ARC value to 0 04 mA MANU DCW VOLTage Set Query Description Sets or returns the DCW voltage in kV The test must first be in DCW mode before this command can be used Note ...

Страница 165: ...U DCW CHISet Set Query Description Sets or returns the DCW HI SET current value in milliamps The test must first be in DCW mode before this command can be used Note A DC Over 50W error will result if the DCW Voltage X HI SET value is 50 watts Note A DC Over 100W error will result if the DCW Voltage X HI SET value is 100 watts IIT 5010 5000 Syntax Query Syntax MANU DCW CHISet NR2 MANU DCW CHISet Pa...

Страница 166: ...se the HI SET range If all the digits in the LO SET range are outside the HI SET range an error will be produced All digits outside the HI SET range are ignored and will not be used For example HI SET value 12 34 LO SET value1 0 005 error LO SET value2 0 053 no error In the example above LO SET value1 will produce an error as all digits are outside the range of HI SET LO SET value2 will not produc...

Страница 167: ...urned off Syntax Query Syntax MANU DCW TTIMe NR2 OFF MANU DCW TTIMe Parameter NR2 OFF 0 5 999 9 seconds TIMER OFF special MANU mode Return parameter NR2 TIME OFF 0 5 999 9 seconds TIMER is OFF special MANU mode Example MANU DCW TTIM 1 Sets the DCW test time to 1 second MANU DCW REF Set Query Description Sets or returns the DCW reference value in mA The test must first be in DCW mode before this co...

Страница 168: ...rrent Set Query Description Sets or returns the DCW ARC current value in mA ARC must be enabled to set the ARC current The test must first be in DCW mode before this command can be used ARC current uses the same range as the HI SET value The ARC current is limited to 2X the HI SET value Syntax Query Syntax MANU DCW ARCCurrent NR2 MANU DCW ARCCurrent Parameter Return parameter NR2 1 000 20 00 IIT 2...

Страница 169: ... Syntax MANU IR VOLTage NR2 MANU IR VOLTage Parameter Return parameter NR2 0 05 1 0 05kV to 1kV steps of 05 IIT 5010 also includes a 0 125kV point Example MANU IR VOLT 1 Sets the IR voltage to 1 kV MANU IR RHISet Set Query Description Sets or returns the IR HI SET resistance value in MΩ IIT 2010 or GΩ IIT 5010 The test must first be in IR mode before this command can be used Syntax Query Syntax MA...

Страница 170: ...alue to Example IIT 2000 2010 MANU IR RHIS 10 Sets the IR HI SET resistance to 10 MΩ Example IIT 5000 5010 MANU IR RHIS 0 010 Sets the IR HI SET resistance to 10 MΩ MANU IR RLOSet Set Query Description Sets or returns the IR LO SET resistance value in MΩ IIT 2010 or GΩ IIT 5010 The LO SET value must be less than the HI SET value The test must first be in IR mode before this command can be used Syn...

Страница 171: ...0 2010 MANU IR RLOS 10 Sets the IR LO SET resistance to 10MΩ Example IIT 5000 5010 MANU IR RLOS 0 010 Sets the IR LO SET resistance to 10MΩ MANU IR TTIMe Set Query Description Sets or returns the IR test time in seconds The test must first be in IR mode before this command can be used Syntax Query Syntax MANU IR TTIMe NR2 MANU IR TTIMe Parameter Return parameter NR2 1 0 999 9 seconds Example MANU ...

Страница 172: ...fore this command can be used The reference value must be lower than the HI SET value Syntax Query Syntax MANU IR REF NR1 MANU IR REF Parameter Return parameter NR1 IIT 2010 only 0000 9999 unit MΩ IIT 5010 only Format A 0 50 00 unit GΩ Format B 0G 50 00G Format C 0M 50000M Example IIT 2000 2010 MANU IR REF 900 Sets the IR reference to 900 MΩ Example IIT 5000 5010 MANU IR REF 0 900 Sets the IR refe...

Страница 173: ... NR2 MANU GB CURRent Parameter Return parameter NR2 3 00 33 00 IIT 2010 3 00 33 00 IIT 5010 Example MANU GB CURR 3 00 Sets the GB current to 3 00A MANU GB RHISet Set Query Description Sets or returns the GB HI SET resistance value in mΩ The test must first be in GB mode before this command can be used Syntax Query Syntax MANU GB RHISet NR2 MANU GB RHISet Parameter Return parameter NR2 000 1 650 0 ...

Страница 174: ... SET value must be less than the HI SET value The test must first be in GB mode before this command can be used Syntax Query Syntax MANU GB RLOSet NR2 MANU IR RLOSet Parameter Return parameter NR2 0 000 649 9 Example MANU GB RLOS 50 Sets the GB LO SET resistance to 50mΩ MANU GB TTIMe Set Query Description Sets or returns the GB test time in seconds The test must first be in GB mode before this com...

Страница 175: ...GB test frequency in Hz The test must first be in GB mode before this command can be used Syntax Query Syntax MANU GB FREQuency 50 60 MANU GB FREQuency Parameter Return parameter 50 50 Hz 60 60 Hz Example MANU GB FREQ 50 Sets the GB test frequency to 50Hz MANU GB REF Set Query Description Sets or returns the GB reference value in mΩ The test must first be in GB mode before this command can be used...

Страница 176: ...be in GB mode and in the Ready Status before this command can be used See page 83 for details on the ZERO function Syntax Query Syntax MANU GB ZEROCHECK ON OFF MANU GB ZEROCHECK Parameter Return parameter ON Zero function is active OFF Zero function is not active Example MANU GB ZEROCHECK OFF Activates the ZERO function MANU UTILity ARCMode Set Query Description Sets or returns the ARC mode status...

Страница 177: ... ARC mode to ON and STOP Example MANU UTIL ARCM OFF Turns ARC mode OFF MANU UTILity PASShold Set Query Description Sets or returns the PASS HOLD setting for the current test Syntax Query Syntax MANU UTILity PASShold ON OFF MANU UTILity PASShold Parameter Return parameter OFF Turns PASS HOLD off ON Turns PASS HOLD on Example MANU UTIL PASS OFF Turns PASS HOLD OFF MANU UTILity FAILmode Set Query Des...

Страница 178: ...e HOLD Sets returns the fail mode as hold STOP Sets returns the fail mode as stop Example MANU UTIL FAIL CONT Sets the fail mode to CONT continue MANU UTILity MAXHold Set Query Description Sets or returns the MAX HOLD setting for the current test Syntax Query Syntax MANU UTILity MAXHold ON OFF MANU UTILity MAXHold Parameter Return parameter OFF Turns MAX HOLD off ON Turns MAX HOLD on Example MANU ...

Страница 179: ...be turned on with the IR and GB function Syntax Query Syntax MANU UTILity GROUNDMODE ON OFF MANU UTILity GROUNDMODE Parameter Return parameter OFF Turns ground mode off ON Turns ground mode on Example MANU UTIL GROUNDMODE ON Turns GROUND MODE on MANU x EDIT SHOW Query Description Returns the test parameters of a manual test Query Syntax MANU x EDIT SHOW Parameter Return parameter x NR1 000 100 Man...

Страница 180: ...15 Version No 001 string Returns a string in the following format Test function test voltage HI SET value LO SET value Ramp time test time Example MANU1 EDIT SHOW ACW 0 100kV H 01 00mA L 00 00mA R 000 1S T 001 0S Returns the test parameters of manual test number 1 ...

Страница 181: ...eep mode the voltage and current settings and the number data points that are used in the last sweep There can be a maximum of 190 data points depending on the testing time The data is returned as a string in the following format SWEEP MODE VSET ISET Get Data data points Query Syntax SWEEP DATA STATus Return parameter string SWEEP MODE VSET unit ISET units Get Data number of data points Example SW...

Страница 182: ...oints are evenly distributed There can be up to 190 data points If only the data from a single point is returned then the data is returned in the following format DATA POINT VSET ISET TIME CR LF If the all the data for the all the points is returned then the data is returned in the following format ACW MODE CR LF No V kV I mA T S CR LF 001 0 071 0 032 0000 1 CR LF 002 0 111 0 047 0000 2 CR LF 013 ...

Страница 183: ...rns the data at point 10 which is at the 1 second time for the sweep test All Data Points Example SWEEP0 DATA SHOW ACW MODE CR LF No V kV I mA T S CR LF 001 0 071 0 032 0000 1 CR LF 002 0 111 0 047 0000 2 CR LF 013 0 601 0 215 0001 3 CR LF END This will return all the data from the sweep graph SWEEP GRAPh SHOW Set Query Description Turns the sweep graph on or off on the IIT 5010 5000 display Synta...

Страница 184: ...Sets or returns which lines are shown on the sweep graph Syntax Query Syntax SWEEP GRAPh LINE NR1 SWEEP GRAPh LINE Parameter Return parameter NR1 Description 0 Turn all lines off all lines are off 1 Displays the graph line for the primary test item See page 87 for details For example V for ACW DCW and IR tests I for GB tests 2 Displays the graph line for the secondary test items For example I for ...

Страница 185: ...or returns the start time STA t of the sweep graph in milliseconds This setting will also set what the time will be for the first point for the sweep data that is returned in the SWEEP DATA SHOW query Syntax Query Syntax SWEEP STARt TIME NR2 SWEEP STARt TIME Parameter Return parameter NR2 0 1 1999 8 seconds Example SWEEP STARt TIME 1000 0 Sets the sweep start time to 1000 seconds ...

Страница 186: ...x PAGE SHOW 185 AUTO PAGE MOVE 185 AUTO PAGE SWAP 186 AUTO PAGE SKIP 187 AUTO PAGE DEL 187 AUTO NAME 188 AUTO EDIT ADD 188 TESTok RETurn 189 SRE 190 AUTO STEP Set Query Description Sets or queries the AUTO number automatic test number Syntax Query Syntax AUTO STEP NR1 AUTO STEP Parameter Return parameter NR1 1 100 ...

Страница 187: ...st in the following format step1 MANU number step2 MANU number step3 etc Query Syntax AUTO x PAGE SHOW Parameter x NR1 1 100 Example AUTO1 PAGE SHOW 01 011 02 004 03 003 04 014 05 015 06 020 07 012 08 018 09 10 11 12 13 14 15 16 Shows the Page View for AUTO number 1 AUTO PAGE MOVE Set Query Description Moves the source step to the desired destination Syntax AUTO PAGE MOVE Value1 Value2 Parameter V...

Страница 188: ...8 0 8 0 0 5 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 0 0 m A E S 0 T _ I AUTO PAGE SWAP Set Query Description Swaps the source step with destination step Syntax AUTO PAGE SWAP Value1 Value2 Parameter Vaue1 NR1 1 16 source step Value2 NR1 1 16 destination step Example AUTO PAGE SWAP 1 4 Swaps the contents of step 1 with step 4 A CW 0 0 0 k 1 O V E M WA P S K I P S D E L E M A U O T N A _ 0 A U T O 0 0 1 0...

Страница 189: ... 0 0 1 0 1 M A N U _ N M E A V H 0 1 0 1 0 0 2 0 0 1 0 3 0 0 3 0 4 0 0 4 0 5 0 0 7 0 6 0 0 3 0 7 0 3 8 0 8 0 0 5 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 0 0 m A E S 0 T _ I AUTO PAGE DEL Set Query Description Deletes the selected step from the AUTO test The remaining steps move up to replace the deleted step Syntax AUTO PAGE DEL NR1 Parameter NR1 1 16 step no Example AUTO PAGE DEL 3 Deletes the contents...

Страница 190: ...ote only alphanumeric characters A Z a z 0 9 and the _ underscore character can be used to set the AUTO test name Syntax Query Syntax AUTO NAME string AUTO NAME Parameter Return parameter string 10 character string first character must be a letter Example AUTO NAME program1 Sets the AUTO name to program1 AUTO EDIT ADD Set Query Description Add the selected MANU test to the current AUTO number Synt...

Страница 191: ... 5 0 0 7 0 6 0 7 0 8 0 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 MANU test added to last step 1 0 0 m A E S 0 T _ I TESTok RETurn Set Query Description Allows OK to be displayed on the remote terminal when a test has stopped PASS FAIL or STOP This applies for MANU and AUTO mode By default TESTok RETurn is set to OFF Syntax Query Syntax TESTok RETurn ON OFF TESTok RETurn Parameter Return parameter ON Enables t...

Страница 192: ...measurement step number at the current point in time during AUTO MODE testing Example User send command SRE to IIT 20X0 50X0 during AUTO mode Query Syntax SRE Return parameter NR1 00 16 Example SRE 5 The current test step is number 5 This indicates that steps 1 4 have already been completed and the results for those steps can now be retrieved ...

Страница 193: ...the internal registers Syntax CLS IDN Query Description Queries the model number serial number and firmware version of the tester Query Syntax IDN Return parameter string Returns the instrument identification as a string in the following format IIT 2000 XXXXXXXXXXXX V1 00 Model number IIT 2000 Serial number 12 character serial number Firmware version V1 00 ...

Страница 194: ...hen this command is used RMT will no longer be displayed on the front panel indicating that remote mode has been terminated Syntax RMTOFF Special Functions Inter Lock Key Open 192 Inter Lock Key Open Set Description This special function is not a command When in remote mode the IIT 20X0 50X0 will send the message Inter Lock Key Open if a test is started with INTERLOCK set to ON but the ...

Страница 195: ...08 2015 Version No 001 193 interlock signal I O pins are not shorted either with the included interlock key or manually This special function is analogous to the INTERLOCK OPEN message that is displayed on the front panel under the same conditions page 119 ...

Страница 196: ...d The possible error messages returned from SYST ERR query are listed below Error Error Code Command Error 0x14 Value Setting Error 0x15 String Setting Error 0x16 Query Error 0x17 MODE Setting Error 0x18 Time Error 0x19 DC Over 50W IIT 2000 2010 only 0x1A DC Over 100W IIT 5010 5000 only 0x1A GBV 5 4V 0x1B ...

Страница 197: ...l not start testing The accuracy does not match the specification The tester will not turn on Ensure the power cord is connected Ensure the line input is set to the correct line voltage Check to make sure the fuse is not blown See page 197 The panel keys are not working Ensure the tester is not in remote mode page 134 Ensure the tester is not in SIGNAL I O or Remote Connect mode page 117 ...

Страница 198: ... testing If Interlock is enabled the interlock key must be inserted into the signal I O port on the rear before a test can be started See page 127 for details Lastly ensure that the Start Ctrl setting is correctly configured in the Common Utility menu For example to enable the START button to start a test ensure that the Start Ctrl setting is set to FRONT PANEL See page 117 for details The accurac...

Страница 199: ...8 2015 Version No 001 197 APPENDIX Fuse Replacement Steps 1 Turn the instrument off POWER 2 Remove the power cord 220 1 0 0 120 230 3 Remove the fuse socket using a flat screwdriver 220 1 0 0 120 230 4 Replace the fuse in the fuse holder 220 1 0 0 1 20 230 ...

Страница 200: ...sert the fuse socket 220 1 0 0 120 230 220 1 0 0 120 230 Rating The fuse for the IIT 2000 2010 and the IIT 5010 5000 have different ratings IIT 2000 2010 100V 120V T5A 250V 220V 230V T2 5A 250V IIT 5010 5000 100V 120V T10A 250V 220V 230V T6 3A 250V Test Errors The following error messages or messages may appear on the IIT screen when configuring or running tests ...

Страница 201: ...W tests OVER 50W is displayed if the HI SET setting multiplied by the Voltage setting is greater than 50W IIT 2000 2010 only OVER 100W IIT 5010 5000 For DCW tests OVER 100W is displayed if the HI SET setting multiplied by the Voltage setting is greater than 100W IIT 5010 5000 only I ERR For ACW DCW tests Shown when the current is set too high SHORT Voltage is too low or there is no High Voltage ou...

Страница 202: ...GB tests The resistance is too high I SET For GB tests Current too low Indicates that the SOURCE L or SOURCE H test lead is open or poorly connected Test the test lead connection with the DUT to confirm I SET For GB tests Current is too high R 0 For GB tests Resistance 0 This error indicates that there is an error with the measured resistance 0Ω Perform the zeroing function again GBV OVER GBV 5 4V...

Страница 203: ...nsation Storage 10 C 70 C 85 No condensation Installation Location Indoors at an amplitude of up to 2000m AC Withstanding Voltage Output Voltage Range 0 100kV 5 000kV Output Voltage Resolution 2V Output Voltage Accuracy 1 of setting 5V with no load Maximum Rated Load 200 VA 5kV 40mA IIT 2000 2010 500 VA 5kV 100mA IIT 5010 5000 Maximum Rated Current 40mA IIT 2000 2010 100mA IIT 5010 5000 0 001mA 10...

Страница 204: ...010 1uA 0 001mA 0 001mA 0 999mA 0 01mA 01 00mA 09 99mA 0 1mA 010 0 040 0mA IIT 5000 5010 1uA 0 001mA 0 001mA 1 100mA 0 01mA 01 11mA 11 00mA 0 1mA 011 1 100 0mA Current Measurement Accuracy IIT 2000 2010 1 5 of rdg 30 counts when HI SET 1 00mA 1 5 of rdg 3 counts when HI SET 1 00mA IIT 5000 5010 1 5 of rdg 30 counts when HI SET 1 11mA 1 5 of rdg 3 counts when HI SET 1 11mA Judgment Valid Range ACW ...

Страница 205: ...U mode MANU 000 DC Withstanding Voltage Output Voltage Range 0 100kV 6 000kV Output Voltage Resolution 2V Output Voltage Accuracy 1 of setting 5V with no load Maximum Rated Load 50W 5kV 10mA IIT 2000 2010 100W 5kV 20mA IIT 5010 5000 Maximum Rated Current 10mA IIT 2000 2010 20mA IIT 5010 5000 0 001mA 2mA 0 100kV V 0 5kV 0 001mA 10mA 0 5kV V 6kV IIT 2000 2010 0 001mA 20mA 0 5kV V 6kV IIT 5000 5010 V...

Страница 206: ...surement Accuracy IIT 2000 2010 1 5 of rdg 30 counts when HI SET 1 00mA 1 5 of rdg 3 counts when HI SET 1 00mA IIT 5010 5000 1 5 of rdg 30 counts when HI SET 1 11mA 1 5 of rdg 3 counts when HI SET 1 11mA Judgment Valid Range DCW HI SET 0 011mA to 1 100mA 00 11mA to 10 00mA IIT 5010 5000 11 00mA 001 1mA to 020 0mA IIT 5010 5000 only LOW SET 0 010mA to 1 099mA 00 10mA to 09 99mA IIT 5010 5000 10 99m...

Страница 207: ... Output Voltage Accuracy 1 of setting 5V with no load Resistance Measurement Range 1M 9500M IIT 2010 1M 50G IIT 5010 Test Voltage Measurement Range Accuracy IIT 2010 50V V 450V 1 50MΩ 5 of reading 1 count 51 2000MΩ 10 of reading 1 count 500V V 1000V 1 500MΩ 5 of reading 1 count 501 9500MΩ 10 of reading 1 count IIT 5010 50V V 450V 0 001 0 050GΩ 5 of reading 1 count 0 051 2 000GΩ 10 of reading 1 cou...

Страница 208: ... Output Current Range 03 00A 30 00A IIT 2010 03 00A 32 00A IIT 5010 Output Current Accuracy 1 of setting 0 2A when 3A I 8A 1 of setting 0 05A when 8A I 30A IIT 2010 1 of setting 0 05A when 8A I 32A IIT 5010 Output Current Resolution 0 01A Frequency 50Hz 60Hz selectable Ohmmeter Measurement Accuracy 1 of reading 2m Ohmmeter Measurement Range 10m 650 0m depending on output current IIT 2010 ...

Страница 209: ... Measurement Resolution 0 1m Windows Comparator Method Yes TIMER Test Time 0 5S 999 9S GND OFF Interface REMOTE Remote terminal Yes SIGNAL IO Yes RS232 Yes USB Device Yes GPIB Yes OPTION General DISPLAY 240 x 64 dot matrix LED back light LCD MEMORY AUTO MANU mode 100 memory blocks total POWER SOURCE AC100V 120V 220V 230V 10 50Hz 60Hz ...

Страница 210: ...0 W x 148 H x 482 D mm Max 24kg Max Table 1a Output Limitation in Withstanding Voltage Testing IIT 2000 2010 Upper Current Pause Output Time AC 30mA I 40mA At least as long as the output time Maximum 240 seconds 0 001mA I 30 mA Not necessary Continuous output possible DC 0 001mA I 10 mA Not necessary Continuous output possible GB 15A I 30A At least as long as the output time 999 9 3A I 15A Not nec...

Страница 211: ...t least as long as the output time Maximum 240 seconds 0 001mA I 80 mA Not necessary Continuous output possible DC 0 001mA I 20 mA Not necessary Continuous output possible GB 15A I 32A IIT 5010 At least as long as the output time 999 9 3A I 15A Not necessary 999 9 NOTE Output Time Ramp Time Test Time ...

Страница 212: ...15 Version No 001 IIT 2000 Dimensions 134 0 POWER START STOP REMOTE MANU AUTO EDIT SAVE UTILITY ESC PAGE PASS FAIL READY TEST Tester HIGH VOLTAGE CAUTION 5 0 kVAC MAX 6 0 kVDC MAX AC DC Withstanding Voltage Insulation Resistance RETURN 329 7 322 0 147 7 452 0 395 9 ...

Страница 213: ...nsions POWER START STOP REMOTE MANU AUTO EDIT SAVE UTILITY ESC PAGE PASS FAIL READY TEST HIGH VOLTAGE CAUTION 5 0 kVAC MAX 6 0 kVDC MAX 329 7 322 0 134 0 147 7 458 7 395 9 Tester SENSE H SOURCE H SENSE L SOURCE L RETURN AC DC Withstanding Voltage Insulation Resistance Ground Bond GB Rx HI POT IR ...

Страница 214: ...imensions 322 0 134 0 147 7 530 9 593 7 Tester AC DC Withstanding Voltage Insulation Resistance Ground Bond PASS FAIL TEST READY CAUTION HIGH VOLTAGE 5 0 kVAC MAX 6 0 kVDC MAX HI POT IR RETURN SENSE L SOURCE L GB Rx SENSE H SOURCE H UTILITY EDIT SAVE MANU AUTO REMOTE STOP START POWER ESC PAGE ...

Страница 215: ...5 Version No 001 213 IIT 5000 Dimensions POWER START STOP REMOTE ESC PAGE HIGH VOLTAGE CAUTION 5 0 kVAC MAX 6 0 kVDC MAX RETURN PASS FAIL READY TEST MANU AUTO EDIT SAVE UTILITY 322 0 134 0 147 7 425 9 482 0 AC DC Withstanding Voltage Insulation Resistance Tester ...

Страница 216: ...nd Low Voltage Directive the following standards were applied EMC EN 61326 1 EN 61326 2 1 Electrical equipment for measurement control and laboratory use EMC requirements 2013 Conducted Emission Radiated Emission EN55011 2009 A1 2010 Class A Electrical Fast Transients EN 61000 4 4 2012 Current Harmonics EN 61000 3 2 2014 Surge Immunity EN 61000 4 5 2006 Voltage Fluctuations EN 61000 3 3 2013 Condu...

Страница 217: ...ow Voltage Directive the following standards were applied EMC EN 61326 1 EN 61326 2 1 Electrical equipment for measurement control and laboratory use EMC requirements 2013 Conducted Emission Radiated Emission EN55011 2009 A1 2010 Class A Electrical Fast Transients EN 61000 4 4 2012 Current Harmonics EN 61000 3 2 2006 A2 2009 Surge Immunity EN 61000 4 5 2006 Voltage Fluctuations EN 61000 3 3 2008 C...

Страница 218: ...0 Disposal instructions 6 EN61010 measurement category 4 pollution degree 5 Environment safety instruction 5 External control 121 Interlock key 128 overview 122 remote operation 123 remote terminal 122 signal I O operation 127 signal I O overview 124 FAQ 195 Front panel diagram 15 GPIB installation 26 Ground symbol 4 Interlock key 128 Line voltage selection 24 List of features 11 Manual tests ARC ...

Страница 219: ... Operating precautions 28 Overview 10 Package contents 14 Power on off safety instruction 5 Rear panel diagram 20 Remote control 129 Command list 139 Command syntax 136 function check 133 interface configuration 130 Service operation about disassembly 4 contact 196 Specifications 201 Sweep function 86 Test errors 198 UK power cord 7 Utility settings buzzer 114 Control settings 117 double action 11...

Страница 220: ... neglect accident unauthorized repair alteration contamination or abnormal conditions of operation or handling Any implied warranties arising out of the sale of this product including but not limited to implied warranties of merchantability and fitness for a particular purpose are limited to the above RS Components shall not be liable for loss of use of the instrument or other incidental or conseq...

Страница 221: ...td Suite 23 A C East Sea Business Centre Phase 2 No 618 Yan an Eastern Road Shanghai 200001 China Europe RS Components Ltd PO Box 99 Corby Northants NN17 9RS United Kingdom Japan RS Components Ltd West Tower 12th Floor Yokohama Business Park 134 Godocho Hodogaya Yokohama Kanagawa 240 0005 Japan North America Allied Electronics 7151 Jack Newell Blvd S Fort Worth Texas 76118 U S A South America RS C...

Отзывы: