IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
163
SET value is > 100 watts (IIT-5010/5000).
Syntax
Query Syntax
MANU:DCW:VOLTage <NR2>
MANU:DCW:VOLTage?
Parameter/
Return
parameter
<NR2> 0.100 ~ 6.100 (kV)
Example
MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.
MANU:DCW:CHISet
Set
Query
Description
Sets or returns the DCW HI SET current value in milliamps. The
test must first be in DCW mode before this command can be
used.
Note: A “DC Over 50W” error will result if the DCW Voltage X HI
SET value is > 50 watts.
Note: A “DC Over 100W” error will result if the DCW Voltage X HI
SET value is > 100 watts (IIT-5010/5000)
Syntax
Query Syntax
MANU:DCW:CHISet <NR2>
MANU:DCW:CHISet?
Parameter/
Return
parameter
<NR2> 0.001 ~ 011.0 (IIT-2000/2010)
0.001 ~ 021.0 (IIT-5010/5000)