IIT-2000/2010/5000/5010 Instruction Manual/English
202
05/08/2015 Version No. 001
Voltage Regulation
1% +5V
[Maximum rated load
→ no load]
Voltmeter Accuracy
(1% of 5V)
Current Measurement
Range
0.001mA~040.0mA [IIT-2000/2010]
0.001mA~100.0mA [IIT-5000/5010]
Current Best Resolution
IIT-2000/2010:
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)
IIT-5000/5010:
1uA
0.001mA(0.001mA~1.100mA)
0.01mA(01.11mA~11.00mA)
0.1mA(011.1~100.0mA)
Current Measurement
Accuracy
IIT-2000/2010:
± (1.5% of rdg + 30 counts) when HI SET<1.00mA
± (1.5% of rdg + 3 counts) when HI SET
≥1.00mA
IIT-5000/5010:
± (1.5% of rdg + 30 counts) when HI SET<1.11mA
± (1.5% of rdg + 3 counts) when HI SET≥1.11mA
Judgment Valid Range
(ACW)
HI SET:
0.011mA to 1.100mA
00.11mA to 11.00mA
001.1mA to 040.0mA(IIT-5010/5000 is 100.0mA)