IIT-2000/2010/5000/5010 Instruction Manual/English
168
05/08/2015 Version No. 001
Parameter/
Return
parameter
<NR1> IIT-2010 only:
2 ~ 9999
(unit = MΩ)
IIT-5010 only:
Format A: 0.002 ~ 50.00
(unit = GΩ)
Format B: 0.002G ~ 50.00G
Format C: 2M ~ 50000M
NULL
Sets the HI SET value to
∞.
Example
(IIT-2000/2010)
MANU:IR:RHIS 10
Sets the IR HI SET resistance to
10 MΩ.
Example
(IIT-5000/5010)
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 MΩ.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
MΩ (IIT-2010)
or GΩ (IIT 5010). The LO SET value must be less than the HI
SET value. The test must first be in IR mode before this
command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?