IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
151
Function
ACW, DCW, IR, GB
Judgment
/Status
PASS, FAIL
VIEW
Test voltage
unit
Test current
/Test resistance
unit
resunit
Test time
/Ramp time
T=time+S
R=time+S
Example
(in MANU mode)
MEAS?
>ACW,FAIL,0.024kV,0.013 mA ,R=000.1S
Returns the test result of the current manual test.
Example
(in MANU mode)
MEAS?
>IR,TEST,0.250kV,----Mohm,T=000.2S
Returns ---- as the test result when the reading is invalid.
Example
(in AUTO mode)
MEAS10?
>IR,FAIL,0.250kV,999M ohm,T=010.3S
Returns step 10 of the current automatic result.
MAIN:FUNCtion
Set
Query
Description
Changes the mode between AUTO and MANU.
Syntax
Query Syntax
MAIN:FUNCtion {MANU|AUTO}
MAIN:FUNCtion ?