IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
49
2. Use the scroll wheel to set the test frequency.
ACW, GB 50Hz, 60Hz
Note
The test frequency can only be set for ACW or GB tests.
Setting the Upper and Lower Limits
Background
There is both a LO and HI judgment setting. When the
measured value is below the LO SET setting, the test will be
judged as FAIL. When the value exceeds the HI SET setting
the test will be judged as FAIL. Any measurement between
the LO SET and HI SET setting is judged as PASS. The LO
SET limit cannot be made greater than the HI SET limit.
Steps
1. Press the HI/LO soft-key or use the UP / DOWN
arrow keys to bring the cursor to the HI SET
(ACW/DCW/GB) setting or the LO SET(IR)
setting.