IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
205
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
TIMER (Test Time)
OFF
1
, 0.5S~999.9S
GND
ON/OFF
1
The timer can only be turned off under special MANU mode (MANU=***-000).
Insulation Resistance Test
Output Voltage
50V~1000V *IIT-5010 also includes a 125V test point.
Output Voltage Resolution
50V
Output Voltage Accuracy
(1% of 5V) with no load
Resistance Measurement
Range
1M
~ 9500M
(IIT-2010)
1M
~ 50G
(IIT-5010)
Test Voltage
Measurement
Range
Accuracy
(IIT-2010)
50V≤V≤450V 1~50MΩ
±(5% of r1 count)
51~2000MΩ
±(10% of r1 count)
50
0V≤V≤1000V 1~500MΩ
±(5% of r1 count)
501~9500MΩ
±(10% of r1 count)
(IIT-5010)
50V≤V≤450V 0.001~0.050GΩ
±(5% of r1 count)
0.051~2.000G
Ω
±(10% of r1 count)
50
0V≤V≤1000V 0.001~0.500GΩ
±(5% of r1 count)
0.501~9.999G
Ω
±(10% of r1 count)
10.00~50.00G
Ω
±(20% of r1 count)
Output Impedance
600kΩ
Window Comparator Method Yes