![RS PRO 917-6298 Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/rs-pro/917-6298/917-6298_instruction-manual_1496054031.webp)
IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
29
Do not turn the power off when a test is running, unless in an
emergency.
3. Only use those test leads supplied with the instrument. Leads
with inappropriate gauges can be dangerous to both the
operator and the instrument.
For GB testing, never use the Sense leads on the SOURCE
terminals.
4. Do not short the HIGH VOLTAGE terminal with ground. Doing
so could charge the chassis to dangerously high voltages.
5. Ensure the earth ground of the line voltage is properly
grounded.
6. Only connect the test leads to the HIGH VOLTAGE/SOURCE
H/SENSE H terminals before the start of a test. Keep the test
leads disconnected at all other times.
7. Always press the STOP button when pausing testing.
8. Do not leave the safety tester unattended. Always turn the
power off when leaving the testing area.
9. When remotely controlling the safety tester, ensure adequate
safety measures are in place to prevent: