IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001
127
Using the SIGNAL I/O to Start/Stop Tests
Background
To use the SIGNAL I/O port the Start Ctrl settings have to be
set to SIGNAL I/O in the Common Utility menu.
Panel operation 1. Set the Start Ctrl option to SIGNAL I/O.
Page 117
2. Connect the Input/Output signals to the SIGNAL
I/O port.
3. To start the testing, short the INPUT_STOP and
INPUT_COM line for a minimum of 1ms to put the
tester into READY status.
4. To start the testing, short the INPUT_START and
INPUT_COM lines for a minimum of 1ms.
5. To stop the testing, temporarily short the
INPUT_STOP and INPUT_COM line again.
NOTE
Even if the IIT-2000/2010/5000/5010 is configured to use the
SIGNAL I/O interface, the STOP button on the front panel can
still be used to stop a test.