IIT-2000/2010/5000/5010 Instruction Manual/English
72
05/08/2015 Version No. 001
Running a MANU Test
Background
A test can be run when the tester is in READY status.
Note
The tester cannot start to run a test under the following
conditions:
A protection setting has been tripped; when a protection
setting has been tripped the corresponding error message is
displayed on the screen. See page 198 for a comprehensive
list of the all the setting errors.
The INTERLOCK function is ON and the Interlock key is not
inserted in the signal I/O port (page 117).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START button is pressed
immediately after the STOP button (<0.5s).
Note
When a test is running the voltage output cannot be changed,
unless the test is under the special manual mode. See page
83 for details.
Steps
1. Ensure the tester is in VIEW status for the current
test. Save the current test if necessary.
Page 71