Remote Control Commands
R&S
®
EX-IQ-BOX
312
User Manual 1409.5570.32 ─ 05
9.4.4 CPRI L1 Analysis Subsystem
The L1 analysis (
L1:
) command subsystem contains the remote-control commands for
queryinng the link status.
...............................................................312
.............................................................312
.............................................................312
..............................................................313
...........................................................313
...........................................................313
............................................................314
........................................................314
...........................................................314
......................................................................315
[:SOURce<hw>]:EBOX:CPRI:L1:HDLC:BRATe?
Queries the HDLC bit rate value during link setup.
Return values:
<rate>
float
Example:
EBOX:CPRI:L1:HDLC:BRAT?
Response:
4800
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:L1:RX:LOCal:LOF?
Queries whether the received signal is synchronized correctly at the CPRI breakout
board, i.e. LOF (
L
oss
o
f
F
rame).
Return values:
<lof>
0 | 1 | OFF | ON
Example:
SOUR1:EBOX:CPRI:L1:RX:LOC:LOF?
queris if received data is synchronized.
Response:
1
(error), data lost synchronization.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:L1:RX:LOCal:LOS?
Queries whether LOS (
L
oss
o
f
S
ignal) has occurred in the received signal at the CPRI
breakout board.
Return values:
<lo>
0 | 1 | OFF | ON
CPRI Remote Control Commands
Содержание EX-IQ-BOX