![Rohde & Schwarz EX-IQ-BOX Скачать руководство пользователя страница 317](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982317.webp)
Remote Control Commands
R&S
®
EX-IQ-BOX
302
User Manual 1409.5570.32 ─ 05
Parameters:
<Direction>
OFF | OUTPut | INPut
Deactivate the GPIO interface.
OFF
Deactivates the GPIO interface.
OUTPut
Selects the respective GPIO interface as output, i.e. send a control
signal to the external device.
INPut
Selects the appropriate GPIO interface as input, which means you
receive a signal from the external device, e.g. for trigger purpose.
*RST:
OFF
Example:
SOUR1:EBOX:CPRI:HW:GPIO1:DIR OUTP
outputs the control signal at GPIO port 1.
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:HW:GPIO<ch>:SIGnal
<Signal>
The command determines the type of the control signal. You can define separate signals
for each of the three interfaces.
The numeric suffix
GPIO<ch>
distinguishes between the GPIO interfaces 1, 2, 3 and 4.
CPRI Remote Control Commands