![Rohde & Schwarz EX-IQ-BOX Скачать руководство пользователя страница 293](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982293.webp)
Remote Control Commands
R&S
®
EX-IQ-BOX
278
User Manual 1409.5570.32 ─ 05
Return values:
<Ber>
float
*RST:
0
Example:
SOUR1:EBOX:USER:DIR REC
sets "Receiver" mode.
SOUR1:EBOX:USER:TEST:RX:STAT ON
activates Rx test.
SOUR1:EBOX:USER:TEST:RX:BER?
checks the bit error rate.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:EBOX:USER:TEST:RX:SIGNal
<Signal>
Selects the incoming test signal for testing. The test receiver analyzes both PRBS 16 or
counter test signals.
Parameters:
<Signal>
PRBS | COUNter
PRBS
Selects a PRBS 16 (
P
seudo
R
andom
B
inary
S
equence)
sequence for testing.
COUNTer
Selects a counter test signal.
Range:
PRBS to COUN
*RST:
PRBS
Example:
SOUR1:EBOX:USER:DIR REIC
SOUR1:EBOX:USER:TEST:RX:SIGN COUN
sets to "Receiver" and selects a counter signal for testing.
Manual operation:
See
[:SOURce<hw>]:EBOX:USER:TEST:RX:STATe
<State>
Activates the test receiver.
The R&S
EX-IQ-BOX expects a test signal at the user interface. First, the test receiver
synchronizes to the last received data word, and then generates a test sequence and
compares each generated sample with the corresponding received sample.
Parameters:
<State>
OFF | ON
*RST:
OFF
Example:
SOUR1:EBOX:USER:DIR REC
.
SOUR1:EBOX:USER:TEST:RX:STAT ON
sets the "Receiver" mode and activates Rx test.
User Defined - Remote Control Commands