Electrical
Power supply
Auto-sensing: 100 - 120V / 220 - 240V, 50/60Hz
Power consumption
150VA
Approvals
EN 61010-1:2001, 61010-2-081/A1:2003, 60825-1/A1:2002,
UL 61010-1:2004, IEC 61010-2-081/A1:2003, CAN/CSA-C22.2
No. 61010-1:2004, CAN/CSA-C22.2 No. 61010-2-081:2004,
CAN/CSA-E60825-1:2003
Processing
Control
Software-controlled by external PC
PC operating system
Windows Vista Business (64-bit)
Interface
Ethernet RJ45
Data acquisition
High-speed 16-bit A/D conversion
Samples to be
measured
Fluorescent features on microscope slides within the
specified slide dimensions
Slide magazine
Holds up to 48 slides and allows automatic loading
Environmental
Ambient temperature
Operation: 18 - 28°C (64 - 82°F)
Non-operation: -20 - 60°C (-4 - 140°F)
Relative humidity
Operation: 20 - 80% non-condensing
Storage: 10 - 80% non-condensing
Overvoltage category
II
Pollution degree
2
Method of disposal
Electronic waste (infectious waste)
Physical
Outer dimensions
Width: 60cm (23.6in)
Height: 46.6cm (18.5in)
Depth: 61.8cm (24.4in)
Weight Approximately
38kg
(84.0lb)