4.2.1 Assigning the Same Scan Parameters to Multiple Slides
You can easily assign the same scan parameters to multiple slides.
To assign using the
Apply to slides with parameter set
option button:
Select the
Apply to slides with parameter set
option button in the Parameter
Control (Figure 25) to apply parameter settings to all slide positions with
the same parameter letter code in the Magazine Control.
To assign using the copy and paste functionality:
1.
If open, close the Parameter Control by clicking the
Close Parameter
Control
button ( , Figure 25).
2.
Go to the row of the slide with the preferred parameters in the
Magazine Control, right-click, and select
Copy
from the menu.
3.
Highlight the rows for the slides to process using the same parameters,
right-click, and select
Paste Parameter Settings
from the menu.
To highlight multiple slide rows, hold down the Ctrl key and
click each slide’s row. Or, click a row and then hold down the
Shift key to highlight a range of slides.
If the slide with the preferred parameters contains more than
one cycle, use the copy and paste functionality to specify the
complete slide definition.
To assign using a parameters file:
You can apply the settings from an existing parameter file to the selected
slide(s).
1.
Highlight the slide row in the Magazine Control (Figure 24).
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......