20 |
NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
It is recommended to avoid any external vibrations to the Microarray
Scanner because these can have a negative impact on its operation.
Figure 12: Dimensions of the Microarray Scanner (Top View and Side View)
Power Requirements
Power supply of the Microarray Scanner is auto-sensing and able to operate
without any manual adjustments within the following voltage ranges:
100 - 120V (AC), 50/60Hz, 150VA
220 - 240V (AC), 50/60Hz, 150VA
3.4 Environmental Requirements
The Microarray Scanner is intended for indoor operation under controlled
laboratory conditions. For optimal operation and smooth running of the
relevant applications, operate the Microarray Scanner only in the specified
conditions. Care should be taken to maintain a low-dust, low-vibration
environment. Keep liquids and vapors away from the Microarray Scanner.