D Maintenance and Care
The maintenance and care procedures are important to prolong the
instrument's life and reduce the need for servicing. This chapter contains
information about the following:
ILC test
Liquid spill cleanup
Ozone filter replacement
Instrument relocation
Instrument disinfection
1. ILC Test
STOP
If this is the first time you are running the ILC test, refer to page 27
for more detailed instruction.
Roche NimbleGen recommends you perform an ILC test every 2
weeks.
1.
Switch on the Microarray Scanner.
2.
Start the instrument software.
3.
Click the
QC
tab and then
ILC
in the instrument software. The
Integrated Laser Check window opens (Figure 16).
4.
Click the
Start/stop ILC procedure
button ( ).
After the ILC test is completed, the instrument software produces and
displays a comprehensive test report containing all measured
parameters, evaluation limits, and results (PASS or FAIL).
For more information on the checks that occur during the ILC test,
refer to “ILC Module” on page 4.