NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
8.
Specify rows and columns by clicking the arrow buttons according to
the array layout. Columns and rows are separated by intersecting lines
that give rise to multiple scan area segments wherein additional single
scan areas are then displayed.
9.
Make sure that the height and width of your initial scan area enables
displaying multiple scan areas within the predefined magenta feature
area.
10.
If necessary, clear the
Link to Geometry
checkbox:
By default, the
Link to Geometry
checkbox is selected. If you resize the
initial scan area, the instrument software automatically scales all other
areas to the same dimension.
Clear the
Link to Geometry
checkbox to disable this functionality. In
each corner and side of a scan area, handles will appear to adjust each
scan area individually.
11.
Click
Apply
to confirm your settings.
12.
If desired, click
Save
to save settings to an area file for future use.
To use an existing array image as template for your area layout:
From the Application menu, select
Open Image
and import the image from
the file directory to the Image View. Then click
Show Image
to display the
image in the Area Definition View.
Alternatively, click the
Display Images
button ( ) in the Magazine Control
in your current session or open a session to retrieve an image. Then click
Show Image
to display the image in the Area Definition View.
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......