NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
| 27
3.8 Performing System Verification: The Initial ILC Test
After you install the Microarray Scanner, control unit, and instrument
software, perform an initial ILC test.
After the initial test, Roche NimbleGen recommends you perform
the ILC test every 2 weeks.
STOP
Running the ILC test requires the ILC slide be installed in the
Microarray Scanner. You will find the ILC slide in its protective
package in the slide magazine box. Wear powder-free gloves or
equivalent hand coverings when handling the ILC slide. Avoid
touching or damaging the featured slide surface
.
1.
If necessary, switch on the Microarray Scanner using the main power
switch on its rear and the on/off button on its left side.
You will notice that the internal fans switch on. The blue
initialize/function light indicates that the instrument is ready for
operation.
2.
Start the instrument software to connect to the Microarray Scanner.
The Microarray Scanner performs an initialization routine.
3.
Click the
QC
tab in the instrument software and then click
ILC
.
Figure 15: QC tab on the Instrument Software Workspace
The Integrated Laser Check window opens (Figure 16) and displays all
available test points. The following figure identifies the components of
the Integrated Laser Check window.
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......