NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
| 21
Temperature
Do not expose or locate the Microarray Scanner near direct sunlight or heat
sources to avoid problems with misaligned mechanics and/or optics. To
assure optimal performance, it is recommended to operate the Microarray
Scanner in a temperature controlled (air conditioned) environment with a
constant temperature.
Temperature for operation: 18°C (64°F) - 28°C (82°F)
Temperature for transport: -20°C (-4°F) - 60°C (140°F)
Humidity
The Microarray Scanner is potentially sensitive to condensation under
humid conditions. Allow time for thermal equilibrium before opening the
shipping box or after moving to a place with changed environmental
conditions.
Humidity for operation: 20 - 80% non-condensing
Humidity for storage: 10 - 80% non-condensing
3.5 Installation of the Control Unit
To install the control unit and connect it to the Microarray Scanner:
1.
Connect the mouse, keyboard, and monitor to the computer as
instructed in the computer’s documentation.
2.
Connect the Microarray Scanner to the computer’s PCI-card Ethernet
port using the Ethernet cable provided with the Microarray Scanner.
3.
Connect the computer to your local network using the main-board
Ethernet port on the back of the computer.
4.
Connect the computer, monitor, and Microarray Scanner to a clean
uninterruptable power supply.
3.6 Installation of the Instrument Software
Following is information to be aware of before you install the instrument
software and details the installation procedures.
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......