68 |
NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
2. Liquid Spill Cleanup
Always switch off the instrument before removing any kind of spills.
All spills must be treated as potentially infectious. Therefore, always
adhere to applicable safety precautions (including the wearing of
powder-free gloves, safety glasses, and protective clothing) to avoid
potential infectious disease contamination.
All resulting waste from the cleanup must be treated as potentially
infectious and the disposal must be performed according to the
information given in
Chapter A, Overview
. If the spill occurs inside of
the instrument, a service technician is required.
1.
Switch off the Microarray Scanner.
2.
Wipe up the spill immediately using absorbent material.
3.
Dispose of contaminated material appropriately.
4.
Clean the instrument surfaces with a mild detergent.
5.
Disinfect the instrument as described in the “Instrument Disinfection”
section on page 70.
6.
Wipe dry all cleaned areas.
3. Ozone Filter Replacement
The need for replacing the ozone filter units may vary with the usage of the
microarray scanner. To ensure optimal performance, replacement of the
ozone filter units will be included in the annual maintenance by your Roche
Field Service Engineer or is available upon request.
4. Instrument Relocation
Before you move the instrument to a different location, remember to:
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......