70 |
NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
5. Instrument Disinfection
STOP
Authorized trained personnel, wearing disposable gloves, protective
glasses, and protective clothing, should perform the disinfection
procedure in a well-ventilated room. Be aware that the disinfectant
can influence instrument performance if you apply it to the inside of
the instrument.
Risk of fire and explosion. Alcohols, such as ethanol or isopropanol,
are flammable and when improperly handled can lead to explosions
and/or fire. Proper laboratory safety precautions must be observed.
The disinfection procedure should be performed according to
national, regional, and local regulations.
Perform the disinfection procedure when cleaning up a spill and
when relocating the instrument.
1.
Disconnect the instrument from the main power supply.
2.
Disconnect the instrument from any accessories.
3.
Carefully wipe all the outside surfaces of the instrument with a clean
lint-free cloth that has been soaked in the disinfecting solution, such as:
•
Lysetol (Schülke & Mayr, Norderstedt/Germany,
www.schuelke-mayr.com
)
•
Aseptisol (Bode Chemie, Hamburg/Germany,
www.bode-chemie.com
)
•
70% ethanol, if neither of the above are available
4.
Perform the same disinfection procedure on the slide magazine.
5.
Repeat the disinfection procedure on any accessories.
6.
Complete the Certificate of Decontamination (refer to page 71) and
affix to the outside of the package.
Содержание Nimblegen MS 200
Страница 1: ......
Страница 2: ......
Страница 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Страница 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Страница 103: ......