9. COMMUNICATION DATA DESCRIPTION
IMS01T28-E4
9-95
9.3.3 Setting parameters exclusive to Z-TIO-G module
Sensor bias
RKC communication
identifier
JI
Modbus
register address
ch1: 0422H (1058)
ch2: 0423H (1059)
This is used to make a bias setting of the RTD input.
Attribute:
R/W
Digits:
7 digits
Number of data:
2 (Data of each channel)
Data range:
3.0000 to
3.0000
Factory set value:
0.0000
Related parameters:
Input type (P. 9-45)
This parameter is active when the instrument has RTD input.
PFF power supply frequency
RKC communication
identifier
JJ
Modbus
register address
0424H (1060)
Set power supply frequency of the Power feed forward (PFF) input.
Attribute:
R/W
Digits:
7 digits
Number of data:
1 (Data of each module)
Data range:
0: 50 Hz
1: 60 Hz
2: Automatic setting
Factory set value:
2
Related parameters:
Power feed forward (PFF) selection (P. 9-96),
Measurement power supply frequency (P. 9-97)
This parameter is activated when the Power feed forward (PFF) input is used.
The SRZ should be used with the default value “2: Automatic setting” in normal operations.
Содержание Z-TIO-G
Страница 8: ...i 6 IMS01T28 E4 MEMO...
Страница 16: ...IMS01T28 E4 1 8 MEMO...
Страница 22: ...2 6 IMS01T28 E4 MEMO...
Страница 34: ...4 MOUNTING 4 4 IMS01T28 E4 4 2 Dimensions Terminal type module 85 100 5 Unit mm 30 6 7...
Страница 52: ...5 14 IMS01T28 E4 MEMO...
Страница 82: ...7 18 IMS01T28 E4 MEMO...
Страница 211: ...TROUBLE SHOOTING IMS01T28 E4 10 1 10 1 Solutions for Problems 10 2...
Страница 216: ...10 6 IMS01T28 E4 MEMO...
Страница 217: ...SPECIFICATIONS IMS01T28 E4 11 1...
Страница 230: ...11 14 IMS01T28 E4 MEMO...
Страница 244: ...A 4 IMS01T28 E4 MEMO...
Страница 245: ...The first edition SEP 2012 IMQ00 The fourth edition FEB 2019 IMQ00...