9. COMMUNICATION DATA DESCRIPTION
IMS01T28-E4
9-5
Operation mode state monitor
RKC communication
identifier
L0
Modbus
register address
ch1: 0008H (8)
ch2: 0009H (9)
Each operation mode state of the Z-TIO module is expressed in bit data items.
Attribute:
RO
Digits:
7
digits
Number of data:
2 (Data of each channel)
Data range:
RKC communication:
ASCII code data
The operation mode state is assigned as a digit image in ASCII code data of 7 digits.
ASCII code data of 7 digits:
Most significant digit ············ Least significant digit
Data:
0: OFF 1: ON
Modbus:
0 to 15 (bit data)
The operation mode state is assigned as a bit image in binary numbers.
Bit image: 0000000000000000
Bit 15 ····················· Bit 0
Bit data: 0: OFF 1: ON
Factory set value:
Related parameters:
Auto/Manual transfer (P. 9-15), Remote/Local transfer (P. 9-16),
RUN/STOP transfer (P. 9-16), Operation mode (P. 9-37)
When the
Operation mode (P. 9-37)
is “0: Unused,” all operation mode state monitor data is
“0: OFF.”
Bit 0:
Control STOP
Bit 1:
Control RUN
Bit 2:
Manual mode
Bit 3:
Remote mode
Bit 4 to Bit 15: Unused
Least significant digit: Control STOP
2nd digit:
Control RUN
3rd digit:
Manual mode
4th digit:
Remote mode
5th digit to Most significant digit:
Unused
Содержание Z-TIO-G
Страница 8: ...i 6 IMS01T28 E4 MEMO...
Страница 16: ...IMS01T28 E4 1 8 MEMO...
Страница 22: ...2 6 IMS01T28 E4 MEMO...
Страница 34: ...4 MOUNTING 4 4 IMS01T28 E4 4 2 Dimensions Terminal type module 85 100 5 Unit mm 30 6 7...
Страница 52: ...5 14 IMS01T28 E4 MEMO...
Страница 82: ...7 18 IMS01T28 E4 MEMO...
Страница 211: ...TROUBLE SHOOTING IMS01T28 E4 10 1 10 1 Solutions for Problems 10 2...
Страница 216: ...10 6 IMS01T28 E4 MEMO...
Страница 217: ...SPECIFICATIONS IMS01T28 E4 11 1...
Страница 230: ...11 14 IMS01T28 E4 MEMO...
Страница 244: ...A 4 IMS01T28 E4 MEMO...
Страница 245: ...The first edition SEP 2012 IMQ00 The fourth edition FEB 2019 IMQ00...