Bridging
if additional termination resistance is
used to
bridge
onto live circuits with minimal
interruption.
Termination Resistance
lists the resistance of
the dummy load connected to the line.
Test Duration
allows you to set the length of
time to run the test from a list of values. If
UFED Control option is enabled, the maximum
value is limited to15 min.
Impulse Separation Time
(IST) specifies the
time interval between successive impulses.
Some impulse events comprise many sub-
pulses or a pulse-train. Use a larger IST to
capture longer events of many sub-pulses, or a
smaller IST between events to capture fast
sub-pulses as separate impulse events.
Threshold Settings
button opens a new page
that allows you to set the
Power Detection
Level
and
WB Impulse Noise
thresholds.
Result
displays the
Count, Detection Level,
Elapsed Time
, and pass/fail status which is
periodically refreshed, until you stop the test.
Test Status
displays the current state of
measurement or an error message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
WB Impulse Noise Threshold Settings
The
Threshold Settings
page allows you to set the
WB Impulse Noise Threshold
values for the test.
Power Detection Level
defines the threshold
limits.
Counter
allows you to set the threshold value
for this counter.
Type
is the threshold limit for each pair: either
Maximum
or
Minimum
.
Enabled
is a flag to use the current threshold:
either
Yes
or
No
.
To set threshold values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Histogram
The
Histogram
tab records the
Elapsed Time
, the
Total Counts
of impulse durations that are collected
during each plot interval, and pass/fail status. The bar
graph displays the impulse counts vs time.