6100-Gfast Multifunction Network Analyzer Operation Manual
120
© 2020 Radiodetection Ltd
Single Tone
The
Single Tone
tab allows you to configure the
following parameters:
Current Group
is the current test group name,
for example
Frequency Tests
.
Current Test
displays the test being
measured. Here you can switch to another
Frequency test from the given list. However,
this action will terminate the ongoing
measurement test.
Receive Tone
allows you to select either a
VF
(voice frequency) or
WB
(wideband) signal.
WB
selection is only available when the
30
MHz Wideband Testing
software option is
installed.
Termination Resistance
defines how much
resistance the tester uses to connect to the
line, and is set as follows:
o
for
VF Tone
, a read-only value of
600
Ω.
o
for
WB Tone
, select 100, 120, or
135
Ω.
FED Frequency
sets the frequency that the
TS125 FED will transmit and is a value in Hz
for
VF Receive Tone
(default 300 Hz) and kHz
for
WB
(default 196 kHz).
Tone Duration
modifies the duration value of
the tone from 5 - 60 seconds. Default is
5 s
.
Result
displays the values for both the RX
VF/WB Loss and Frequency
Test Status
displays the current state of
measurement or an error message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the
desired parameter.
2. Press the left/right arrow keys to view and
select the options.
OR
3. Press
on a value to open a list box of
options or the alphanumeric editor screen and
use the navigation keys to scroll through.
4. Press
to confirm the value.
Single Tone Result Details
The
Result Details
read-only page displays the saved
receive tone or
VF/WB RX Loss/Frequency
values for
the
Single RX Tone with FED
test.